Chapter 6
First Light
45
6.2.1.1
Rotational Alignment
Perform the following procedure to rotationally align the system:
1. Click on the View tab located just above Experiment Devices to change to the
.
Figure 6-10: Typical Spectroscopy Application Workspace: View Tab
2. If you have not already done so, mount a light source (e.g., Neon, Argon, or Xenon,)
to the front of the entrance slit.
3. Open the Spectrometer expander, select the grating, and select the center
wavelength.
4. Set the slit to 10
m. If necessary, adjust the Exposure Time to maintain optimum
(i.e., near full-scale,) signal intensity.
5. Wait until the camera temperature locks at its default temperature.
6. Verify that the spectroscopy-mount adapter moves freely at the spectrograph.
NOTE:
It may be necessary to loosen the two set screws that lock
the sliding tube in place.
44
11
-0
16
9_0
0
2
2
Summary of Contents for Princeton Instruments NIRvana HS
Page 1: ...www princetoninstruments com NIRvana HS System Manual 4411 0169 Issue 1 0 January 8 2020 ...
Page 3: ...Issue 1 0 NIRvana HS System Manual 3 This page is intentionally blank ...
Page 4: ...4 NIRvana HS System Manual Issue 1 0 ...
Page 10: ...10 NIRvana HS System Manual Issue 1 0 ...
Page 26: ...26 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Page 34: ...34 NIRvana HS System Manual Issue 1 0 ...
Page 52: ...52 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Page 72: ...72 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Page 92: ...92 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Page 101: ...This page is intentionally blank ...
Page 106: ......