CAUTION!
!
Never set the coolant temperature below the dew point.
Coolant temperature should not be below the dew point of the ambient air. Internal
condensation caused by operation below the dew point may damage the camera and
will void the warranty. Additionally, the Lansis monitors its internal temperature and will
automatically restart its fan regardless of the software settings if it senses excessive
internal heat buildup.
2.1.4.3
Coolant Ports
The liquid cooled model of the Lansis camera is equipped with cooling ports that allow
it to be connected to an external chiller. As is the case with circulating air, circulating
coolant removes heat produced by the camera. This means of heat removal is designed
for vibration-free data acquisition. For the circulating coolant to function properly, free
air circulation must be maintained between the sides of the external circulator and the
laboratory atmosphere.
CAUTION!
!
NEVER apply negative pressure to the liquid circulator
fittings on the Lansis camera. Doing so may permanently
damage the camera.
CAUTION!
!
The wetted areas of the Lansis heat exchanger and fittings
are nickel plated.