Portable Emissions Analyzer
Appendix
Teledyne
Analytical
Instruments
59
Appendix
Specifications
Sensors:
O
2
Measurement _____________________________
Sensor: Electrochemical, cell
Range: 0-20.9 vol.% O
2
Resolution: 0.1 vol. %
Accuracy: ± 0.2% vol.
Response Time (T97): <70 sec
CO2 Determination _____________________________
(Calculated from O2)
Sensor:
none—calculated
value
Display
Range:
0-CO
2max
vol. %
Resolution: 0.1 vol. %
Accuracy:
0.2%
vol.
Response time (T97): <70 sec
CO
Measurement
_____________________________
(with H2 compensation)
Sensor:
Electrochemical,
cell
Measurement Range: 0-4000 ppm
Resolution:
1
ppm
Accuracy: ± 5 ppm (up to 150 ppm)
± 5% of reading (up to 4000 ppm)
Response Time (T90): <60 sec
NO
Measurement
_____________________________
(Option)
Sensor:
Electrochemical,
electrochemical
cell
Measurement Range 0-2000 ppm
Resolution:
1
ppm