Operator’s Manual
27
HVD3000 / HVD3000A Test Record
Serial Number:____________________________________________________
Asset / Tracking Number: _________________________________________
Date:______________________________________________________________
Technician:________________________________________________________
Equipment
Model
Serial Number
Calibration Due Date
Oscilloscope
Digital Multimeter
Function Generator*
* In this Performance Verification Procedure, the function generator is used for making relative measurements. The
output of the generator is measured with a DMM or oscilloscope. Thus, the generator is not required to be calibrated.
Step
Description
Intermediate Data
Test Result
5
Function Generator Output Voltage
V
11
Probe Output Voltage
V
13
Gain Error (test limit
± 1 %)
%
16
Probe Output Voltage
V
18
Gain Error (test limit
± 1 %)
%
Permission is granted to photocopy this page to record the results of the Performance Verification procedure.
File the completed record as required by applicable internal quality procedures.
•
Line numbers correspond to steps in the procedure that require the recording of data.
•
Record the actual specification limit check under "Test Result". The test limits are included in all of
these steps.
•
Record other measurements and intermediate calculations that support the limit check under
"Intermediate Results".
Summary of Contents for HVD3000
Page 1: ...Operator s Manual HVD3000 HVD3000A High Voltage Differential Probes...
Page 2: ......
Page 3: ...HVD3000 HVD3000A High Voltage Differential Probes Operator s Manual November 2018...
Page 16: ...HVD3000 HVD3000A High Voltage Differential Probes 10 HVD3106 6M HVD3106A 6M TYPICAL BANDWIDTH...
Page 47: ......
Page 48: ...930490 00 Rev A November 2018...