If the system pressure is higher than 250 psig (1.7 MPa) the pressure induced error in the span reading
becomes significant. This error will approach 16% for nitrogen at 1000 psig. For accurate high-pressure
measurements this error must be corrected.
The formula for predicting the nitrogen mean error expressed as a fraction of the reading are:
Sensor Type
A
B
0.026”
-2.2915*10
-7
5.7198*10
-5
0.017”
-7.1066*10
-8
5.2403*10
-5
0.014”
-5.3091*10
-8
6.1278*10
-5
𝐸𝑟𝑟𝑜𝑟
= 𝐴𝑃 + 𝐵𝑃
Where P is the pressure in psig and
Error
is the fraction of the reading in error.
The flow reading can be corrected as follows:
𝐶𝑜𝑟𝑟𝑒𝑐𝑡𝑒𝑑 = 𝐼𝑛𝑑𝑖𝑐𝑎𝑡𝑖𝑜𝑛 − (𝐼𝑛𝑑𝑖𝑐𝑎𝑡𝑖𝑜𝑛 × 𝐸𝑟𝑟𝑜𝑟)
Where the
Indication
is the indicated flow and
Error
is the result of the previous formula (or read from
charts above).