Teledyne HASTINGS 4/6 Manual Download Page 6

148-0598 Rev. A 

Page 6 

May  1998 

vacuum.  Just install the gauge tube in 
the vacuum system. 
 

To change the current configuration:

 

 

3.2.8

  With the AVC module still 

plugged into either a DV-4 or DV-6 tube 
and the tube still at atmospheric 
pressure, turn the MODEL 4/6 

OFF

 

3.2.9

  Press and hold the push button 

and turn the MODEL 4/6 

ON

.  As soon 

as the power is ON, release the push 
button.  The display will read 8888 
 

3.2.10

  Press and release the push button 

within 5 sec.  The display will read the 
program code.  For example “A_2.0” or 
higher. 
 

3.2.11

  Press and release the push button 

again within 5 sec.  The display will read 
“rSEt”. Press and release again within 5 
sec. to reset the original factory 
calibration. (If you have made a custom 
calibration, do nothing and the display 
will advance to the next mode in about 5 
sec. leaving your custom calibration in 
memory.) 
 

3.2.12

  When the program has advanced 

to next mode, the display will read either 
“torr” or “hPA” (millibar).  If the 
pressure units are correct do nothing and 
the display will automatically advance to 
the next mode in about 5 sec.  If you 
wish to change the pressure units, press 
and release the push button within 5 sec.  
The new pressure units will be displayed 
and in about 5 sec. the display will 
advance to the next program mode. 
 

3.2.13

  When the program has advanced 

to the next mode, the display will read 
either “d-4”, “d-6”, d-4C”, or “d-6C” 
(NOTE:  The d-4 and d-6 settings are not 

used with the AVC module.)  If the tube 
setting is correct, do nothing and the 
display will automatically advance to the 
next mode in about 5 sec.  If you wish to 
change the tube setting, press and release 
the push button within 5 sec. until the 
correct tube setting is displayed.  In 
about 5 sec. the display will advance to 
the next program mode. 
 

DO NOT PRESS THE MODE 
SWITCH AGAIN UNTIL THE 
MODEL 4/6 RETURNS TO THE 
MEASUREMENT MODE AND 
READS  “_or_”.  

This takes about 50 seconds. 
 

NOTE

:  The remaining modes, 

CAL.H 

and 

CAL.L

 are for calibration and could 

change the calibration if the tube is at 
atmospheric pressure. 
 

3.3 INSTALLING 

VACUUM 

 GAUGE 

TUBE

 

 
The Vacuum Gauge Tube is a 

required 

accessory

 because of the wide variety of 

tubes that can be used. 
 

3.3.1 

The vacuum gauge tube is 

shipped with a protective cap covering 
the evacuation port to keep the 
thermopile elements clean and free from 
physical damage.  Remove the protective 
cap prior to installing the tube. 
 
 

3.3.2 

The vacuum gauge tube can be 

mounted in any position without 
affecting calibration.  The recommended 
position is with the tube vertical and the 
stem down so that condensable vapors 
do not collect in the tube. 
 

3.3.3 

Most tubes will have a 1/8” NPT 

threaded stem.  This should be screwed 

Summary of Contents for HASTINGS 4/6

Page 1: ...Page 1 May 1998 HASTINGS MOD 4 6 DIGITAL VACUUM INSTRUMENT TELEDYNE BROWN ENGINEERING HASTINGS INSTRUMENTS P O Box 1436 Hampton VA 23661 USA Telephone 757 723 6531 800 950 2468 FAX 757 723 3925 May 1998 148 0598 Rev A ...

Page 2: ...NTIFY 2 3 0 INSTRUMENT SET UP TUBE INSTALLATION 3 3 1 WIRING 3 3 2 CONFIGURATION 3 3 3 INSTALLING VACUUM GAUGE TUBE 4 4 0 OPERATION 5 4 1 AVC MODULE 5 4 2 TAKING A VACUUM MEASUREMENT 5 4 3 RESOLUTION OF DIGITAL METER 5 5 0 CALIBRATION 5 5 1 RESTORE FACTORY CALIBRATION 5 5 2 CUSTOM CALIBRATION 6 5 3 REFERENCE TUBE CALIBRATION 7 6 0 MAINTENANCE 8 7 0 TROUBLE SHOOTING AND REPAIR 8 7 1 TROUBLE SHOOTIN...

Page 3: ...ch heats a pair of noble metal thermocouples A change in pressure in the gauge tube changes the thermal conductivity of the gas surrounding the thermocouples thus changing the heat loss of the thermo couples due to thermal conductivity This results in a temperature change in the thermocouples thus changing their dc millivolt output The dc millivolt output varies inversely with pressure with 10mv e...

Page 4: ...r use Hastings Instruments shall not be liable for any indirect or consequential damages Teledyne Brown Engineering Hastings Instruments will at its option repair replace or refund the selling price of the product if Teledyne Brown Engineering Hastings Instruments determines in good faith that it is defective in materials or workmanship during the warranty period provided that the product is retur...

Page 5: ...d d 6C To check your current configuration 3 2 1 Plug the AVC module into either a DV 4 or DV 6 tube The tube should be at atmospheric pressure 3 2 2 Remove the front bezel and the front panel Locate the black push button in the lower right hand corner 3 2 3 Turn the MODEL 4 6 ON or plug into 115vac Since the tube is at atmospheric pressure the display should read _or_ for over range It may read E...

Page 6: ...splay will advance to the next program mode 3 2 13 When the program has advanced to the next mode the display will read either d 4 d 6 d 4C or d 6C NOTE The d 4 and d 6 settings are not used with the AVC module If the tube setting is correct do nothing and the display will automatically advance to the next mode in about 5 sec If you wish to change the tube setting press and release the push button...

Page 7: ...LUTION OF DIGITAL METER 4 3 1 The resolution of the digital meter varies with pressure For the DV 4 0 10 to 20 Torr range the resolution is Range Resolution 0 010 to 0 100 Torr 0 002 Torr 0 10 to 1 00 Torr 0 01 Torr 1 0 to 10 0 Torr 0 1 Torr 10 to 20 Torr 1 Torr 4 4 3 The resolution of the digital meter varies with pressure For the 0 001 to 2 Torr range the resolution is Range Resolution 0 001 to ...

Page 8: ... wish to change the tube setting press and release the push button within 5 sec until the correct tube setting is displayed In about 5 sec the display will advance to the next program mode DO NOT PRESS THE MODE SWITCH AGAIN UNTIL THE MODEL 4 6 RETURNS TO THE MEASUREMENT MODE AND READS _or_ This takes about 50 seconds 5 2 CUSTOM CALIBRATION CAUTION The MODEL 4 6 will only accept two high calibratio...

Page 9: ...e DB 16D which is used with the DV 4 range or the DB 20 is which is used with the DV 6 range The DB 16D should have a marked calibration value between 0 01 Torr and 0 20 Torr The DB 20 should have a marked calibration value between 0 001 Torr and 0 010 Torr 1 to 10 millitorr You do not need access to the push button for this calibration procedure 5 3 1 Connect the Reference Tube to the AVC module ...

Page 10: ...on when tube was not at the correct pressure REMEDY Re configure MODEL 4 6 or activate Factory Reset 7 2 WARRANTY REPAIR Hastings Instruments will repair or replace at their option any vacuum product covered under the warranty expressed in Section 1 5 that is returned transportation prepaid to their plant in Hampton Virginia All products returned must be accompanied by the Service Form or a writte...