FIRE
R M LD
FIRE
R M LD
Operator’s Manual
11
GAS LASER
Doc No. 16162 Rev A, GAS LASER Operator Manual
© 2021 Teledyne Gas Measurement Instruments Ltd. All rights reserved.
Battery Pack
Dual Bay Battery Charger
- Provides the power to the
instrument
- Removable
- Rechargeable
- Lithium-ion
- Up to 8 hours of operating time
when fully charged.
- Charge two batteries
simultaneously
- Includes a universal AC adapter
and power cord
- An integrated metal bracket
included on the bottom of the
charger allows for wall or fixed
mounting options.
To wall mount charger use four 8-32 screws to thread into self-clinching nuts
on charger chassis. Thread dept is 0.060in, max screw intrusion is .150in.
NOTICE: Access to both sides of the mounting wall are needed.
NOTICE: For wall mounted or mobile applications, Teledyne recommends
securing the battery to the charger using the captive screws attached to
the battery pack.
NOTICE
Two (2)
charging
bays
Universal AC adapter
Power cord
WARNING: To reduce the risk of
ignition of a flammable or explosive
atmosphere, battery pack must be
removed, inserted and recharged
only in a location known to be
non-hazardous.
WARNING: Only use the Teledyne
supplied Gas Laser battery pack
charger to recharge the battery
pack. Use of any other charger
may cause severe damage to the
battery pack or electrical circuits.
Battery Charger is not part of the
hazardous safety certification.
AVERTISSEMENT
WARNING
AVERTISSEMENT : Pour réduire
le risque d’inflammation d’une
atmosphère inflammable ou
explosive, bloc-batteries doivent
être retirées, insérées et rechargées
uniquement dans un endroit réputé
non dangereux.
AVERTISSEMENT : Utiliser
uniquement le chargeur de
bloc-batterie Gas Laser fourni par
Teledyne pour recharger le bloc-
batterie. L’utilisation de tout autre
chargeur peut endommager
gravement le bloc-batterie ou
les circuits électriques. Le chargeur
de batterie ne fait pas partie de la
certification de sécurité dangere-
use.
NOTICE
NOTICE: Must recharge between
uses to assure no interruption in
use.
NOTICE: The prolonged storage of
the battery pack inside or outside
the instrument can lead to battery
chemistry being irreversibly
damaged leading to permanent
failure of the battery pack.
NOTICE: Do not dispose of this
product in the unsorted municipal
waste stream. Dispose of this
product according to
local regulations.