Falcon GC
Maintenance & Troubleshooting
Teledyne Analytical Instruments
25
if the flame is lit by holding a reflective surface over
the FID exhaust and examining for condensation.
Check with a flow meter to ensure that the Air (ignite)
Pressure, the Air (Run) Pressure and the Hydrogen
Pressure values are consistent with those found in your
Data Package shipped with the instrument.
5.2 Adjusting FID Pressures
To adjust the pressure of in the FID please follow the steps described
below.
Note: For Ambient and N
2
Carrier Gas applications, the settings
below may be different. Refer to the Tech Note provided
with the specific instrument.
1.
Turn off the H
2
and air flow to the FID. If carrier gas is N
2
adjust
the flow to 5psi.
2.
Click “Remote” on the Chromperfect screen.
3.
Select the “Pressures” tab.