48
Figure 6-4: Micrometer Locking Set Screw
17.
Click
Run
, and, while continuously acquiring data, adjust the micrometer until:
•
The intensity level has been maximized, and
•
The FWHM for a selected peak or peaks has been minimized.
NOTE:
The
Peak Find
function for the graph selected by the cursor is
automatically turned on to assist in identifying peaks and viewing
FWHM widths.
shows before and after alignment graphs.
Figure 6-5: Typical Alignment Helper Views: Rotational Alignment
18.
Once properly aligned, tighten the micrometer locking set screw.
19.
Place the Micrometer Cover on the spectrograph.
20.
Replace and tighten the four (4) cover screws.
21.
Stop data acquisition.
44
11-0
143_0
015
L
OCKING
S
ET
S
CREW
441
1-01
43_00
31
B
EFORE
A
LIGNMENT
A
FTER
A
LIGNMENT