GT-FIRE
PORTABLE GAS DETECTOR
67816
Revision D
44
7.2. Viewing Bump Test Results
From CalGas set-up screen (shown in Figure
53: Example CalGas Cylinder
), press and hold LAST
button to view the results of the last 32 bump tests.
The screen shown in
Figure 57: Previously Stored Bump Tests
will be displayed.
Figure 57: Previously Stored Bump Tests
• Use the UP
or DOWN
buttons to highlight the required results.
• Press and hold the NEXT
button to select.
Once selected, the stored bump test reading will be shown (as illustrated in Figure
58: Stored Bump
Test Readings
).
Figure 58: Stored Bump Test Readings
To return to CalGas Display, press and hold the OK
button.
7.3. Bump Test Logging
All manual bump tests results will be automatically stored in the monitor’s memory when the mode
is exited using the EXIT
button.
Up to 32 bump tests are stored in chronological order. When more bump tests are performed, the
oldest result will be over-written.
Results can be extracted from the monitor using the GT Data Downloading software (part no.
67164).