User Manual
CP Series
Version 3.1 – June 2020
67
Command description
Fields description
Command syntax
Description
Reply from PowerBox
Field name
Description
units
#TP<XXXX>
Program new exposure time
value
!TP<YYYY>,<E>,<AAAA>
<XXXX>
Requested exposure time value
s
<YYYY>
Programmed exposure time value
<E>
0 : no error
N.A.
1 : no generator connected
2 : generator rental date is over
3 shot is in progress
4: requested value is below min TP value
5: requested value is above max TP value
6: requested TP is bigger than remaining rental
time
7: preheat is required – TP is set automatically
<AAAA>
If <E> = 0,1,2,3,6
0
N.A.
If <E> = 4
min TP value for generator
s
If <E> = 5
max TP value for generator
s
#ON
Start shot
!ON
N.A.
N.A.
N.A.
#OF
Stop shot
!OF
N.A.
N.A.
N.A.
#XO
Watchdog
!XO
N.A.
N.A.
N.A.
?TS
Ask for the shot type
!TS<X>
<X>
0: None (no shot actually)
N.A.
1: Preheating
2: Normal shot
3: Cycles
4: Conditioning
?SN
Ask for the serial numbers of
generator and PowerBox
!SN<XXXXXXXX>,<YYYYYY
YY>
<XXXXXXXX>
Serial number of generator
N.A.
<YYYYYYYY>
Serial number of PowerBox
N.A.
?MD
Ask for the generator model
!MD<XXXXXXXXX>
<XXXXXXXXX>
Generator model
N.A.