10.4.6
Auto-adjust region
By using the
Auto-adjust region
feature it is possible to select an area of the image, and
then have the colorization of the image adjusted based on the temperatures within that
selected area.
To configure an auto-adjust region, follow this procedure:
1. On the toolbar and in the
Data presentation modes
section, click
Auto-adjust region
.
2. In the image, click and then draw a region of a desired size.
3. To move the region, click and hold the region label in the image, and then move it into
position.
4. To resize the region, click and hold one of the corners of the region in the image, and
then move it to the desired size and shape.
5. To remove the region, click
X
in the region label.
10.4.7
Calibrating the camera
The calibration of the camera is performed as a non-uniformity correction (NUC). A NUC
is
an image correction carried out by the camera software to compensate for different
sensitivities of detector elements and other optical and geometrical disturbances
1
.
Calibration is needed whenever the output image becomes spatially noisy. This may, for
example, happen when the ambient temperature changes. The calibration can be carried
out automatically with different selectable intervals. It is also possible to perform a cali-
bration manually.
10.4.7.1
Automatic calibration
To set the automatic calibration, follow this procedure:
1. On the toolbar and in the
Calibration tools
section, click
Calibration
. This displays a
list of calibration options.
2. Click one of the options:
Option
Calibration is automatically carried out:
Auto
when needed.
Every 10 min
approximately every 10 minutes, or when needed.
Every 30 min
approximately every 30 minutes, or when needed.
Every 60 min
approximately every 60 minutes, or when needed.
Every [type your value]
at any set interval.
OFF
never (automatic calibration disabled).
3. While the automatic calibration is in progress, the
Notification
text
Calibration in prog-
ress ...
is temporarily displayed under the live image.
10.4.7.2
Manual calibration
To perform a manual calibration, follow this procedure:
1. On the toolbar and in the
Calibration tools
section, click
Calibration
. This displays a
list of calibration options.
2. Click
Calibrate now
.
3. While the manual calibration is in progress, the
Notification
text
Calibration in prog-
ress ...
is temporarily displayed under the image on the screen