8800 Series Trace Moisture Analyzer Instruction Manual
30
Legend:
+
denotes a beeping accompaniment to the message.
/
denotes alternately flashing messages.
DISPLAY
EXPLANATION
REQUIRED ACTION
LO
ERR
+
PROM check sum failed.
power
on tes
ts
HI
ERR
+
RAM write/read test failed.
cycle power
%
ERR
+
Unidentified power-up failure.
if problem persists,
C.S.F/
+
EEPROM
C
heck
S
um
F
ailed.
return to your representative
ERR/AD
+
A/D
converter failure.
sy
st
em
te
st
s
on
ce p
er 2
m
in
.
for service.
ERR/REF
+
Ref
erence voltage for A/D out of spec.
LO/BAT
+
Lo
w
supply
voltage.
LO
RNG/
+
TEMP
Instrument
lo
w
temp
erature
rang
e
has been exceeded.
make sure that the unit is at a
HI
RNG/
+
TEMP
Instrument
hi
gh
temp
erature
rang
e
has been exceeded.
temperature of -10°C to +50°C.
OPN/
+
Sensor circuit is
open
.
dewpoint dis
play
te
st
s
SAT/
+
Sensor is
sat
urated.
see table in sensor troubleshooting section 2.5
SHR/
+
Sensor circuit is
shor
ted.
...
Trying to calculate dewpoint for undefined sensor.
select sensor and recalibrate.
DB
+
Trying to calibrate an undefined sensor.
cal
ib
ra
tio
n
te
st
s
see calibration instructions.
SEN/2LO
+
Sen
sor reading is ‘
too’ lo
w to be from a saturated sensor,
thus can not be calibrated
2LO
While Compensating for cable length, the measurement did not
correspond to the expected open or short values.
cabl
e
te
st
s
Check the means of opening & shorting the sensor
cable. see section 3.4.5 - 11
2HI
ERR/EEP
+
EEP
ROM write cycle not completed.
m
is
cel
laneous
if this persists, return for service.
LOC
+
Attempting to modify a
loc
ked unit.
unlock unit, see set-up state 3.4.5.-10.
SEC
+
Attempting to unlock a
sec
ured unit.
HI
(flashing while
viewing DEWPOINT)
The HI Alarm (alarm #1) relay is deenergized.
See section 3.4.4.2 Setting the Alarms.
LO
(flashing while
viewing DEWPOINT)
The LO Alarm (alarm #2) relay is deenergized.
A.O./OPN
+
The Analog Output ‘sees’ an open circuit
check the cables connecting the analog output, make
sure the terminating impedance is less that 500
Ω.
PSI
flashing when dis-
playing dewpoint
Pressure correction is enabled, the dewpoint is computed using
programmed pressures
refer to section 3.4.4.1
-70
°C (
-94
°F) with
flashing
DEWPOINT
An XTR-65 sensor is measuring less than -70°C or -94°F D.P.
check instrument
upgrade to XTR-100 sensor
RNG
number can not be displayed in 3½ digits
XEN
+
Turn on message,
Xen
(
Greek
Ξ
=X)
display is dark
No power to instrument
check power to instrument, check fuse