06160J DCN7969
Teledyne API Model 465M O
3
Monitor User Manual
39
ENT
SPAN CAL
900
ENT
ENT TO
SPAN
SUCCESSFUL?
ENT
RETURN TO
CONC MODE
Yes
OUT OF
RANGE
No
ENT
Enter Span target
concentration (actual ozone
concentration being supplied
to monitor)
CFG
Figure 7-4: Span Cal Menu
7.2.4.
VIEW Menu
The VIEW menu allows the operator to view various measurement parameters in real-
time. This can be useful for diagnosing various instrument or system problems.
Table 7-5: VIEW Menu Parameters
Parameter
Meaning
Units
Normal Range*
MEAS
UV detector reading, measure cycle
mV
250 – 1230
REF
UV detector reading, reference cycle
mV
250 – 1230
PRESSURE Sample pressure
psia
9.0 – 16.0
LAMP
TEMP
UV lamp temperature
ºC
50.0 – 52.0
SAMPLE
TEMP
Sample temperature
ºC
5.0 –50.0
FLOW
Sample Gas Flow Rate**
ON/OFF
ON
*After initial 20 minute warm-up period.
** Optional Hardware.
NOTE
These values are updated in real-time as they are repeatedly
scrolled on the display.