06160J DCN7969
Teledyne API Model 465M O
3
Monitor User Manual
11
2.
SPECIFICATIONS AND AGENCY
APPROVALS
2.1.
Model 465M Specifications
Measuring Principle
UV absorption (Beer Lambert Law)
Ranges
User selectable to any full scale range from:
0-1000 PPM to 0-10,000 PPM
0-5 g/Nm
3
to 0-50 g/Nm
3
0-1 wt% to 0-3 wt%
Measurement Units
PPM, wt%, g/Nm
3
Accuracy
±
1% of Full Scale
Precision/Repeatability
±
0.5% of Full Scale
Display Resolution
1 PPM, 0.001 wt%, 0.01 g/Nm
3
Response Time (95%)
<45 seconds to 95%
Compensation
Pressure, Temperature (NTP = 273.15K, 760
mmHg)
Gas Inlet Pressure Range
11.0 – 16.0 psia
Sample Flow Rate
0.8 LPM
Humidity Range
10-90% RH, non-condensing
Temperature Range
5-45o C
Dimensions (H x W x D)
Rack Mount (3U Panel) - 5.22” (133 mm) x 19.0"
(483 mm) x 15.3" (388 mm)
With NEMA 4X enclosure
, typical product envelope:
16.85" x 15.60 (±0.25)" x 6.65" (428mm x 396 (±6.35)mm
x 169mm)
Weight
Rack Mount – 13.6 lbs (6.17 kg)
With NEMA 4X enclosure– 16.5 lbs (7.5 kg)
Power
100-240 V~, 47-63 Hz, 74 W
Environmental
Installation Category II
Pollution Degree 2
For Indoor Use Only
Maximum Operating Altitude 2000 meters
Analog Output
0 – 5 V or 4-20 mA isolated output
Status Outputs
System OK, Invalid Reading, UV Lamp Fault,
Pneumatic Error
RS-232 (optional RS-485)
57.6 Kbaud, DB-9 connector