Trace Oxygen Analyzer
Installation
Teledyne Analytical Instruments
17
switching up to 0.5 ampere at 125 VAC into a resistive load (2A for 30
VDC).
The alarm relay circuits are designed for failsafe operation,
meaning the relays are energized during normal operation. If power fails
the relays de-energize (alarms activated).
The contact connections are indicated diagrammatically on the rear
panel as Normally Closed, Common, and Normally Open. Figure 3-2
explains how these act in failsafe operation.
Alarm 1 and Alarm 2 can both be configured as either HI or LO.
A HI alarm will activate when concentration is above threshold, while a
LO alarm will activate concentration is below threshold.
Figure 3-2: Contact ID for FAILSAFE Relay Operation
The specific descriptions for each type of alarm are as follows:
Alarm #1
Programmable as high or low alarm (actuates when
concentration is above threshold). Can be set
anywhere between 0 and 9999ppm, but must be set
ABOVE the threshold set for the LO Alarm.
Alarm #2
Programmable as high or low alarm (actuates when
concentration is below threshold). Can be set
anywhere from 0 to 9999ppm.
Sensor Fail
Actuates when the output of the Micro-Fuel Cell
sensor falls below the acceptable level.