2-2: Part II
2 Operational Theory
Model 3010PAC
Teledyne Analytical Instruments
being analyzed. The Micro-Fuel Cell is therefore a hybrid between a
battery and a true fuel cell. (All of the reactants are stored externally in a
true fuel cell.)
2.2.2 Anatomy of a Micro-Fuel Cell
A Micro-Fuel Cell (MFC) is a cylinder only 1¼ inches in diameter
and 1¼ inches thick. It is made of an extremely inert plastic, which can be
placed confidently in practically any environment or sample stream. The
cell is effectively sealed, although one end is permeable to oxygen in the
sample gas. The other end of the cell is a contact plate consisting of two
concentric foil rings. The rings mate with spring-loaded contacts in the
sensor block assembly and provide the electrical connection to the rest of
the analyzer. Figure 2-1 shows the external features of a typical cell.
Figure 2-1: Micro-Fuel Cell
Refer to Figure 2-2, Cross Section of a Micro-Fuel Cell, which illus-
trates the following internal description.
Figure 2-2. Cross Section of a Micro-Fuel Cell (not to scale)
Summary of Contents for 3010PAC
Page 12: ...iv Part I Model 3010PAC Oxygen Analyzer TeledyneAnalyticalInstruments...
Page 18: ...1 6 Part I 1 Introduction Model 3010PAC Teledyne Analytical Instruments...
Page 22: ...2 4 Part I 2 Operational Theory Model 3010PAC Teledyne Analytical Instruments...
Page 34: ...3 Installation Model 3010PAC 3 12 Part I Teledyne Analytical Instruments...
Page 72: ...3 Installation Model 3010PAC 3 6 Part II Teledyne Analytical Instruments...
Page 76: ...4 Maintenance Model 3010PAC 4 4 Part II Teledyne Analytical Instruments...