5 Maintenance
Model 3010
PAC
Oxygen Analyzer
5-2: Part I
Teledyne Analytical Instruments
3. Replace fuse as shown in Figure 5-2.
4. Reassemble Housing as shown in Figure 5-1.
American Fuses
European Fuses
Figure 5-2: Installing Fuses
5.2
System Self Diagnostic Test
1. Press the
System
button to enter the system mode.
2. Use the
< >
arrow keys to move to
More
, and press
Enter
.
3. Use the
< >
arrow keys to move to
Self-Test
, and press
Enter
.
The following failure codes apply:
Table 5-1: Self Test Failure Codes
Power
0
OK
1
5 V Failure
2
15 V Failure
3
Both Failed
Analog
0
OK
1
DAC A (0–1 V Concentration)
2
DAC B (0–1 V Range ID)
3
Both Failed
Preamp
0
OK
1
Zero too high
2
Amplifier output doesn't match test input
3
Both Failed