Operation
3000ZA2G
Teledyne Analytical Instruments
42
4.4.1.2
M
ANUAL
M
ODE
Z
EROING
Press
Zero
to enter the
Zero
function. The screen that appears
allows you to select between automatic and manual zero calibration. Use
the ▲/▼ keys to toggle between
AUTO
and
MAN
zero settling. Stop
when
MAN
appears, blinking, on the display.
Zero: Settling: Man
<ENT> To Begin
Press
Enter
to begin the zero calibration.
After the zero calibration completes its cycle, the information is
stored in the microprocessor, and the instrument automatically returns to
the
Analyze
mode.
4.4.2 Span Cal
The
Span
button on the front panel is used to span calibrate the
analyzer. Span calibration can be performed using the
automatic
mode,
where an internal algorithm compares consecutive readings from the
sensor to determine when the output matches the span gas concentration.
Span calibration can also be performed in
manual
mode, where the
operator determines when the span concentration reading is acceptable
and manually exits the function.
4.4.2.1
A
UTO
M
ODE
S
PANNING
Press
Span
to enter the span function. The screen that appears
allows you to select whether the span calibration is to be performed
automatically or manually. Use the ▲/▼ arrow keys to toggle between
AUTO
and
MAN
span settling. Stop when
AUTO
appears, blinking, on
the display.
Span: Settling: AUTO
<ENT> For Next
Press
Enter
to move to the next screen.