![Teledyne 3000ZA2G Operating Instructions Manual Download Page 21](http://html1.mh-extra.com/html/teledyne/3000za2g/3000za2g_operating-instructions-manual_1080409021.webp)
Introduction
3000ZA2G
Teledyne Analytical Instruments
6
Sample Inlet and Outlet:
One inlet (must be externally
valved) and one exhaust. Three inlets when auto calibration
option is present.
Reference Inlet and Outlet:
Flowing air reference inlet and
exhaust.
9-Pin RS-232 Port:
Serial digital concentration signal
output and control input.
50-Pin Equipment Interface Port:
Analog Outputs:
0–1 VDC oxygen concentration plus
0-1 VDC range ID
Alarm Connections:
2 concentration alarms and 1
system alarm.
Remote Valve
: Used in the 3000ZA2G for controlling
external solenoid valves only.
Remote Span/Zero:
Digital inputs allow external
control of analyzer calibration.
Calibration Contact
: To notify external equipment that
instrument is being calibrated and readings are not
monitoring sample.
Range ID Contacts
: Four separate, dedicated, range
relay contacts. Low, Medium, High, Cal.
Network I/O:
Serial digital communications for local
network access. For future expansion. Not currently
implemented.
Optional
Calibration Gas Ports:
Separate fittings for zero, span, and
sample gas input, internal valves for automatic gas switching.
Current Signal Output:
Isolated 4–20 mA DC plus optional
4–20 mA DC range ID.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 3000ZA2G is accurate to 2-3 %. Accordingly,
internally scheduled calibrations can vary 30-45 minutes
per day.