Operation 3000ZA-3X
Teledyne
Analytical
Instruments
44
4.7.2 Manual Range Select
-----------------------
Man Rng: R1
-----------------------
Pressing ENTER causes the range
R1
or
R2
or
R3
to start flashing.
Then pressing UP or DOWN cycles through the 3 ranges. The selected
range is the effective range if the Range Select (above) is set to
Man
(ual).
4.7.3 Range Setpoints
-----------------------
R1: 0-1000.00 ppm
-----------------------
Pressing ENTER causes the value to the right of
Rn:
to start
flashing. Then pressing UP or DOWN adjusts the value. These are the
range setpoints for
R1
,
R2
, and
R3
. They may be set to a value in the
range of 1.00 ppm to 100% with the stipulation that the each range must
have a setpoint equal to or greater than the setpoint of the range below it
plus 1ppm.