Percent Oxygen Analyzer
Operation 4
4-5
Teledyne Analytical Instruments
4.3.2 Setting up an Auto-Cal
When proper automatic valving is connected (see chapter 3,
installa-
tion
), the Analyzer can cycle itself through a sequence of steps that automati-
cally zero and span the instrument.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 3000MA is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
To setup an
Auto–Cal
cycle:
Choose
System
from the Function buttons. The VFD will display five
subfunctions.
Contrast Auto—Cal
PSWD Logout More
Use
< >
arrows to blink
Auto—Cal
, and press
Enter
. A new screen for
Span/Zero
set appears.
Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h
Press
< >
arrows to blink
Span
(or
Zero
), then press
Enter
again.
(You
won’t be able to set
OFF
to
ON
if a zero interval is entered.)
A
Span
Every
...
(or
Zero Every
...
) screen appears.
Span Every 0 d
Start 0 h from now
Use
∆∆∆∆∆∇
arrows to set an interval value, then use
< >
arrows to move to
the start-time value. Use
∆∆∆∆∆∇
arrows to set a start-time value.
To turn
ON
the
Span
and/or
Zero
cycles (to activate
Auto-Cal
): Press
System
again, choose
Auto—Cal
,
and press
Enter
again. When the
Span/
Zero
screen appears, use the
< >
arrows to blink the
Span
(or
Zero
)
OFF
/
ON
field. Use
∆∆∆∆∆∇
arrows to set the
OFF
/
ON
field to
ON
. You can now
turn these fields
ON
because there is a nonzero span interval defined.
4.3.3 Password Protection
If a password is assigned, then setting the following system parameters
can be done only after the password is entered:
span
and
zero
settings,
alarm
setpoints, analysis
range
definitions, switching between
autoranging
and manual override, setting up an
auto-cal
, and assigning a new
password
.
However, the instrument can still be used for analysis or for initiating a self-
test without entering the password.
Contrast Function is
DISABLED
(Refer to Section 1.6)