Thermal Conductivity Analyzer
Part I: Control Unit
Part I 4-29
Teledyne Analytical Instruments
2.
Use LEFT/RIGHT key to move to the range that is speci-
fied as inverting output.
3.
Press and hold DOWN key for approximately 5 to 7 seconds.
4.
Press ENTER key.
NOTE: If the inverting has been setup, “i” shall display on the left
bottom corner. Otherwise, the left bottom corner display ”n”.
If more that one range is specified an inverting output, repeat steps 1
to 4.
4.9.3 Special - Polarity Coding
NOTE: This setup will be identified only when performing
GAS TEST
or calculation. Formula 1 will determine which range(s)
willrequired polarity coding.
If the analyzer indicates a negative concentration with proper span gas or
formula 1 indicates a negative value, set the S1 accordingly to the table below:
Close S1-5 range 1
Close S1-6 range 2
Close S1-7 range 3
Close S1-8 cal range
Press I/O to restart the system.
4.9.4 Special - Nonlinear Application Gain Preset
NOTE: This section apply during
GAS TEST routine for the unit that
has more than one range install with nonlinear output applica-
tion.
The steps are as follows:
1.
Set unit range to lowest range reading.
2.
Using the computer generated settings for the controller,
adjust the controller settings for the maximum span gas
output.
3.
Press SPAN key. Select AUTO mode and setup the setting
to span level. Press ENTER key to span.
4.
Set range switch to next range.
5.
Press SPAN key.
Summary of Contents for 2010B
Page 16: ...Teledyne Analytical Instruments 1 8 Part I 1 Introduction Model 2010B ...
Page 22: ...2 6 Part I 2 Operational Theory Model 2010B Teledyne Analytical Instruments ...
Page 34: ...Teledyne Analytical Instruments 3 Installation Model 2010B 3 12 Part I ...
Page 92: ...4 Maintenance Model 2010B 4 8 Part II Teledyne Analytical Instruments ...