Thermal Conductivity Analyzer
Part I: Control Unit
Part I 4-5
Teledyne Analytical Instruments
4.3.1 Setting the Display
Contrast Function is
DISABLED
If you cannot read anything on the VFD after first powering up:
1. Observe LED readout.
a. If LED meter reads 8.8.8.8.8., go to step 3.
b. If LED meter displays anything else, go to step 2.
2. Press
I/O
button twice to turn Analyzer OFF and ON again. LED
meter should now read 8.8.8.8.8.. Go to step 3.
4.3.2 Setting up an AUTO-CAL
When proper automatic valving is connected (see chapter 3, installation),
the Analyzer can cycle itself through a sequence of steps that automatically zero
and span the instrument.
Note: Before setting up an AUTO-CAL, be sure you understand the
Zero and Span functions as described in section 4.4, and
follow the precautions given there.
Note: If you require highly accurate AUTO-CAL timing, use external
AUTO-CAL control where possible. The internal clock in the
Model 2010B is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
Note: If all your ranges are for the same gas application, then AUTO-
CAL will calibrate whichever range you are in at the scheduled
time for automatic calibration.
Note: If your ranges are configured for different applications, then
AUTO-CAL will calibrate all of the ranges simultaneously (by
calibrating the Cal Range).
To setup an
AutoCal
cycle:
Choose
System
from the Function buttons. TheVFD will display five
subfunctions.
CONTRAST AUTOCAL
PWD LOGOUT MORE
Contrast Function is
DISABLED
(Refer to Section 1.6)
(Refer to Section 1.6)
Summary of Contents for 2010B
Page 16: ...Teledyne Analytical Instruments 1 8 Part I 1 Introduction Model 2010B ...
Page 22: ...2 6 Part I 2 Operational Theory Model 2010B Teledyne Analytical Instruments ...
Page 34: ...Teledyne Analytical Instruments 3 Installation Model 2010B 3 12 Part I ...
Page 92: ...4 Maintenance Model 2010B 4 8 Part II Teledyne Analytical Instruments ...