4-1
BioOptix™ 10
Section 4 System Maintenance
4.1 Periodic Maintenance
The BioOptix 10 system’s internal components have been
selected for long life and low maintenance. To maintain the
system, the only required periodic maintenance is to flush the
pumps daily. The pump flush passes a wash fluid through the
pumps to prevent salt buffers from crystallizing in the system.
CAUTION
Salts will cause premature wear to the internal components
and eventually equipment failure. Flush the pumps daily to pre-
vent damage.
The BioOptix 10 system provides the option to flush the pump at
the end of a purification run (see section 3.8). If the system will
remain idle for a few hours or even overnight, a pump flush fol-
lowing the run is recommended.
The pump flush can also be initiated from a menu command.
Daily Pump Flush
1. Select
Tools>Edit Pump Flush
and review the settings. Save
any changes you’ve made.
2. Ensure the Buffer In lines are placed in containers of wash
fluid.
Note
Select a wash fluid that is strong enough to dissolve salts. The
wash fluid must also be miscible with buffer solution remaining
in the system from the previous run.
3. Select the
Tools>Pump Flush
menu command. This opens
the BioOptix 10 flush window (Figure 4-1).
4. Check the pump channels to be flushed. Then, click the
Start
button.
Figure 4-1
Flush window
Summary of Contents for 10
Page 2: ......
Page 4: ......
Page 8: ...BioOptix 10 Safety vi ...
Page 36: ...BioOptix 10 Section 2 Installation 2 18 ...
Page 50: ...BioOptix 10 Section 3 Operation 3 14 ...
Page 69: ...BioOptix 10 Appendix B Fluid Path Diagrams B 5 Figure B 7 BioOptix 10 Plumbing Diagram ...
Page 70: ...BioOptix 10 Appendix B Fluid Path Diagrams B 6 ...
Page 72: ...BioOptix 10 Appendix C Replacement Parts Lists C 2 ...
Page 74: ...13BioOptix 10 Replacement Parts Lists C 4 ...
Page 75: ...13BioOptix 10 Replacement Parts Lists C 5 ...
Page 76: ...13BioOptix 10 Replacement Parts Lists C 6 ...
Page 77: ...13BioOptix 10 Replacement Parts Lists C 7 C 2 Column Module ...
Page 78: ...13BioOptix 10 Replacement Parts Lists C 8 ...
Page 79: ...13BioOptix 10 Replacement Parts Lists C 9 ...
Page 80: ...13BioOptix 10 Replacement Parts Lists C 10 ...
Page 81: ...13BioOptix 10 Replacement Parts Lists C 11 C 3 Pump Module ...
Page 82: ...13BioOptix 10 Replacement Parts Lists C 12 ...
Page 83: ...13BioOptix 10 Replacement Parts Lists C 13 ...
Page 84: ...13BioOptix 10 Replacement Parts Lists C 14 ...
Page 86: ......
Page 88: ......