BioOptix™ 10
Section 3 Operation
3-5
Pausing at the end of a block allows for buffer changes or
other mechanical adjustments before proceeding with the
next block.
•
Wavelength
– Type or select the desired wavelength from
190 to 360 nm to be used for UV detection. Note that all
blocks use the same wavelength.
Note
The Equilibrate tab is a “Fluid” block type. See Appendix A.
3.5.2 Load Settings
This block loads the sample onto the column.
Figure 3-3
Load tab settings
The Sample Load tab (Figure 3-3) contains the following:
•
Injection
– Select
Pump
to pump liquid sample from the
Sample In ports onto the column. Select
Manual
to inject
the sample from the manual injection ports (located on
the top of the BioOptix 10) onto the column. Note that
selecting
Manual
disables the flow rate and volume
settings.
•
Flow Rate, mL/min
– If a
Pump
injection is selected, enter
the desired flow rate in mL/min.
•
Volume
– If a
Pump
injection is selected, enter the
desired volume in mL.
•
Collect Effluent
– Select the desired action for effluent
fluid as it comes off the column. If
Yes
is selected, enter
the desired fraction size.
•
Pause at End
– Select
Yes
to have the system wait for
operator intervention after completing this block. Select
No
to automatically advance to the next block.
Note
The Load tab is a “Load” block type. See Appendix A.
Summary of Contents for 10
Page 2: ......
Page 4: ......
Page 8: ...BioOptix 10 Safety vi ...
Page 36: ...BioOptix 10 Section 2 Installation 2 18 ...
Page 50: ...BioOptix 10 Section 3 Operation 3 14 ...
Page 69: ...BioOptix 10 Appendix B Fluid Path Diagrams B 5 Figure B 7 BioOptix 10 Plumbing Diagram ...
Page 70: ...BioOptix 10 Appendix B Fluid Path Diagrams B 6 ...
Page 72: ...BioOptix 10 Appendix C Replacement Parts Lists C 2 ...
Page 74: ...13BioOptix 10 Replacement Parts Lists C 4 ...
Page 75: ...13BioOptix 10 Replacement Parts Lists C 5 ...
Page 76: ...13BioOptix 10 Replacement Parts Lists C 6 ...
Page 77: ...13BioOptix 10 Replacement Parts Lists C 7 C 2 Column Module ...
Page 78: ...13BioOptix 10 Replacement Parts Lists C 8 ...
Page 79: ...13BioOptix 10 Replacement Parts Lists C 9 ...
Page 80: ...13BioOptix 10 Replacement Parts Lists C 10 ...
Page 81: ...13BioOptix 10 Replacement Parts Lists C 11 C 3 Pump Module ...
Page 82: ...13BioOptix 10 Replacement Parts Lists C 12 ...
Page 83: ...13BioOptix 10 Replacement Parts Lists C 13 ...
Page 84: ...13BioOptix 10 Replacement Parts Lists C 14 ...
Page 86: ......
Page 88: ......