Operator’s Manual 

Do not overload. Do not apply any potential to the probe leads that exceeds 

the maximum rating of the probe. Always comply with the Voltage vs. 

Frequency Derating Curve. 
Observe all terminal ratings. Observe all markings on the oscilloscope 

before connecting. Consult the oscilloscope product manual for further 

ratings information. 
Do not remove probe casing. Touching exposed connections may result in 

electric shock. 
Use indoors only within operational environment listed. Do not use in wet or 

explosive atmospheres. Keep product surfaces clean and dry. 
Handle with care. Probe tips are extremely sharp and may puncture skin or 

cause other bodily injury if not handled properly. 
Keep fingers behind the finger guard of probe body and accessories. 
Do not operate with suspected failures. Before each use, inspect the probe 

and accessories for any potential damage such as tears or other defects in 

the probe body, cable jacket, accessories, etc. If any part is damaged, cease 

operation immediately and sequester the probe from inadvertent use. 

Cleaning 

 

Clean only the exterior of the probe, cables, and accessories. Use a 

soft cotton cloth moistened with a mild detergent and water solution. 

Do not allow any portion of the probe to be submerged at any time. 

 

Dry the probe and accessory thoroughly before attempting to make 

any voltage measurement. 

 

Do not subject the probe to solvents or solvent fumes as these can 

cause deterioration of the probe body, cables, and accessories. 

 

Summary of Contents for PP026 Series

Page 1: ...PP026 1 PP026 2 Operator s Manual PP026 Passive Probe...

Page 2: ...roy is a registered trademark of Teledyne LeCroy Inc Other product or brand names are trademarks or requested trademarks of their respective holders Information in this publication supersedes all earl...

Page 3: ...OTECTIVE EARTH TERMINAL Introduction The Teledyne LeCroy PP026 is a passive high impedance voltage oscilloscope probe designed and calibrated for use with general purpose oscilloscopes having an input...

Page 4: ...o 30 C 5 to 75 RH 30 C to 40 C 5 to 50 RH above 40 C RH not controlled below 10 C Operating Altitude 3000 meters maximum To reduce risk of fire or electric shock do not exceed the voltage or category...

Page 5: ...Operator s Manual 3 PP026 Input Impedance Profile Voltage vs Frequency Derating Curve...

Page 6: ...iance until 22 July 2017 per Article 4 Paragraph 3 General Safety Information Observe generally accepted safety procedures in addition to those listed here to avoid personal injury or damage to equipm...

Page 7: ...or cause other bodily injury if not handled properly Keep fingers behind the finger guard of probe body and accessories Do not operate with suspected failures Before each use inspect the probe and acc...

Page 8: ...n oscilloscopes Low Frequency 1 Apply a 1 kHz square wave to the probe or connect to the oscilloscope s calibrator output 2 Adjust the single LF trimmer located on the PP026 BNC Box until you achieve...

Page 9: ...compensated correct and under compensated waveforms Service Refer all service to qualified personnel For the most complete and up to date list of sales and service centers by country visit teledynele...

Page 10: ...PK5 PK6 002 Color Coding Rings 2 ea yellow red green blue 8 5 PK6 5MM 103 Ground Lead with Alligator Clip 1 6 PK6 5MM 104 Ground Spring 1 7 PK6 5MM 105 PCB Adapter 1 8 PK6 5MM 106 BNC Adapter 5mm 1 9...

Page 11: ...oy at its option will either repair or replace any assembly returned within the warranty period to the Customer Service Department of an authorized service center However this will be done only if the...