FAULT DO SPECIFICATIONS
The Fault DO is configured normally low; the output is high when an error has been
detected.
For most error conditions, the Fault DO will continue to indicate a fault until the unit is
successfully re-zeroed or rebooted. However, some errors are inherently intermittent in
nature, and are cleared automatically when they no longer affect instrument data. For
example, the three minute instrument warm-up period is indicated as an error to allow
detection of accidental power disruption; the Fault DO will indicate no fault when the
warm-up period is over.
•
Operating Voltage (Power, pin 8): 24VDC ± 10%
•
Output circuit:
o
Output HIGH Voltage: 24 VDC ± 10%, 203
Ω
output resistance.
o
Output LOW Voltage: 4.9 ± 0.1 VDC, 203
Ω
output resistance.
OVER-VOLTAGE PROTECTION
Maximum Ratings For DB15 Input Lines
•
Forward voltage: 30V
•
Reverse voltage: 50V
DB15 output lines are short-circuit protected.