8
Teledyne API OPTI-Sense 5200
094900000B DCN8470
SPECIFICATIONS
Table 1-1. Specifications
Parameter
Description
Measuring Principle
Non Dispersive IR absorption
Application
Measurement of SiF4 Absorbance
Dimensions (W x H x D)
4.72” x 8.66” x 3.15” (120 mm x 220 mm x 80 mm)
Light Source
Infrared light source
Response Time
(time to respond to step
function of SIF4)
< 2 seconds to 90% of step
Reading Update
1 reading/second
Analog scaling/Filter type
1-10 VDC corresponds to (0.0-0.2) units of absorbance,
< 10Watts, factory set
Leak Testing
Vacuum Leak-Tight Cell
Gas Connection
KF16-300 Series stainless steel
Pressure Range
200mT-5000mT
Temperature Range
5 – 45
°
C
Power Input
15 VDC Customer Supplied, 350 mA max during normal
operation
24VDC Customer Supplied
Weight
7 Lbs. (3.17 kg)
Digital Interface
RS-232, with GUI Interface
Software
Embedded (standard) Labview GUI interface
Compliance
CE approved
FEATURES
•
Zero by Voltage Command
•
Fault Digital Output (DO) indicates instrument errors.
•
1-10 Volt Analog Output
•
Overvoltage Protection
Summary of Contents for OPTI-Sense 5200