background image

 

 

 

 

 

 

User Manual

 

OPTI-Sense 5200 

with 

Fault High, Zero Low, Customer-Supplied Power 

© TELEDYNE API (TAPI) 

9970 CARROLL CANYON ROAD  

SAN DIEGO, CALIFORNIA 92131-1106 

USA

 

Toll-free Phone:  +1 800-324-5190 

Phone:  +1 858-657-9800 

Fax:  +1 858-657-9816 

Email: 

[email protected]

 

Website: 

http://www.teledyne-api.com/ 

 

 

 

Copyright 2022 

094900000B DCN8470 

Teledyne API 

25 March 2022 

 

Summary of Contents for OPTI-Sense 5200

Page 1: ...r TELEDYNE API TAPI 9970 CARROLL CANYON ROAD SAN DIEGO CALIFORNIA 92131 1106 USA Toll free Phone 1 800 324 5190 Phone 1 858 657 9800 Fax 1 858 657 9816 Email api sales teledyne com Website http www te...

Page 2: ......

Page 3: ...OF COPYRIGHT 2022 Teledyne API TAPI All rights reserved TRADEMARKS All trademarks registered trademarks brand names or product names appearing in this document are the property of their respective own...

Page 4: ...ion or proper tools could result in damage to the part s and or the instrument Technician Symbol All operations marked with this symbol are to be performed by qualified maintenance personnel only Elec...

Page 5: ...gnements sp cifiques MISE EN GARDE Surface chaude Ne pas toucher Toucher certaines parties de l instrument sans protection ou sans les outils appropri s pourrait entra ner des dommages aux pi ces ou l...

Page 6: ...vided but not manufactured by TAPI is warranted and will be repaired to the extent and according to the current terms and conditions of the respective equipment manufacturer s warranty PRODUCT RETURN...

Page 7: ...the careful and effective use of the instrument and related pertinent information ATTENTION COULD DAMAGE INSTRUMENT AND VOID WARRANTY This special notice provides information to avoid damage to your i...

Page 8: ...L INSTALLATION 9 3 ELECTRICAL CONNECTIONS 11 RS 232 Connector 11 I O Connector 11 Fault DO Specifications 12 Over Voltage Protection 12 4 SYSTEM OPERATION 13 Zeroing the OPTI Sense 13 5 INSTRUMENT ERR...

Page 9: ...iod is completed CAUTIONS Please review these instructions carefully to ensure proper installation operation and care of the system to minimize the risk of personal injury or product damage due to imp...

Page 10: ...second Analog scaling Filter type 1 10 VDC corresponds to 0 0 0 2 units of absorbance 10Watts factory set Leak Testing Vacuum Leak Tight Cell Gas Connection KF16 300 Series stainless steel Pressure Ra...

Page 11: ...ENTION Ensure adequate space to access connectors and to remove cover when necessary for servicing Ensure adequate clearance to allow free air flow for cooling and to avoid high heat generating equipm...

Page 12: ...is also acceptable Figure 2 2 Inlet right and Outlet left Connectors The flow of gas through the unit should be interrupted when there is a possibility of particulate matter soiling the windows or lo...

Page 13: ...nnected to the COM1 port of the computer running the GUI software I O CONNECTOR Table 3 1 below shows the I O connector pinout Table 3 1 I O Connector Pinout Pin Connection Notes 1 N C 2 Analog Signal...

Page 14: ...automatically when they no longer affect instrument data For example the three minute instrument warm up period is indicated as an error to allow detection of accidental power disruption the Fault DO...

Page 15: ...ow the line must be held low for at least 200 ms Operating Voltage HIGH 24VDC 10 Operating Voltage LOW 0VDC Nominal 4VDC Max Zeroing will take approximately five seconds during normal operation Zeroin...

Page 16: ...r contamination indicates presence of particulate matter or a deposited film Make sure that provisions are in place to prevent particulate matter or film deposition onto the surfaces of the absorption...

Page 17: ...chnical Support department for any technical assistance needed Teledyne API Technical Support 9970 Carroll Canyon Road San Diego California 92131 1106 USA Toll free Phone 1 800 324 5190 Phone 1 858 65...

Reviews: