Functions
3.
Navigate to
Utilities
>
View HW/SW Version
.
4.
Verify that the software version number matches the version of the
software-upgrade package you installed.
SW Installation
Troubleshooting
If a power failure occurs during SW upgrade and the instrument has a blank screen
when you reapply power, perform the following procedure.
1.
Make sure that an Ethernet cable is connected directly from your PC to the
instrument. You need a direct network connection. (See page 5,
2.
Turn off the instrument (remove power).
3.
Press and hold the left arrow key while you press the power button.
4.
The instrument will turn on and the default IP address of 192.168.1.1 will
be set.
5.
Set the IP address of your PC to 192.168.1.x, where x
≠
1.
6.
Enter the IP address of the instrument (192.168.1.1) for the upgrade transfer
application.
7.
Start the upgrade process. When it is complete, the instrument should reboot.
If it does not, cycle the power.
If you continue to experience a problem, contact your Tektronix sales
representative for assistance.
94
WVR5200 Waveform Rasterizer User Manual