Performance Verification
4- 2
WFM90D and WFM91D Service Manual
The tests in this section comprise a valid confirmation of instrument performance
when the following requirements are met:
H
Test equipment used to verify performance requirements must be calibrated
and working within the limits specified in Table 4--2 on page 4--3.
H
The WFM 90D or WFM 91D Handheld Waveform, Vector, Picture, and
Audio Monitor must have last been adjusted at an ambient temperature range
of +20
°
C to +30
°
C, and must have been operating for a warm-up period of
at least 20 minutes.
Summary Verification
The summary verification is listed in Table 4--1 and is intended for those who are
familiar with the complete performance verification procedures. Procedure titles
and page numbers provide a cross-reference to the performance verification
procedures on the following pages.
Table 4- 1: Summary verification procedure
Procedure
Procedure Requirement
Page #
Preliminary setup
Initial equipment connections and control settings.
4--5
Sync separation
Signal synchronization will occur with input signals of composite video or
black burst, with sync amplitudes of 40 IRE (300 mV PAL)
6 dB for
internal reference, and sync amplitudes of between 143 mV and 4 V for
external reference.
4--5
Sweep timing and integral linearity
Sweep Timing Accuracy: 5
s/Div. (1 Line), 10
s/Div. (2 Line), and
1.0
s/Div. (2 Line + MAG),
2%. 0.2
s/Div. (1 Line + MAG)
3%.
Integral Linearity:
1%.
4--6
Vertical gain and vertical magnifier registration
1 Volt Full Scale: 1 V input displayed within 1% of 140 IRE (1.00 V PAL).
X5 Gain: Gain accuracy within 5% with 1 V input signal. X5 Gain
Registration:
≤
1 major division of vertical shift from baseline between
unmagnified and magnified signal.
4--7
Variable gain range and vertical position range
Variable Gain Range: Input signals between 0.8 V and 2 V can be
adjusted to 140 IRE (1.0 V) display. 160 mV and 400 mV for X5 Gain.
Vertical Position Range: 1 V signal can be positioned so that peak white
and sync tip can be placed at blanking level, with the DC Restorer Clamp
on, regardless of gain setting.
4--7
Overscan
≤
2% variation in baseline of 100 IRE (700 mV PAL) 12.5T (20T PAL)
modulated pulse as it is positioned over the middle 80% of the screen.
4--8
Input and DC restorer frequency response
X1 Gain Response (Flat filter selected): 50 kHz to 6 MHz within 2% of
response at 50 kHz. X5 Gain Response (Flat filter selected): 50 kHz to
6 MHz within 5% of response at 50 kHz. Attenuation of 60 Hz (50 Hz PAL)
on Input Signal: SLOW mode:
≤
20%. FAST mode:
≥
90%.
4--9
Prerequisites
Summary of Contents for WFM90D
Page 4: ......
Page 14: ...Service Safety Summary x WFM90D and WFM91D Service Manual...
Page 17: ...Specifications...
Page 18: ......
Page 29: ...Operating Information...
Page 30: ......
Page 44: ...Functional Overview 2 14 WFM90D and WFM91D Service Manual...
Page 57: ...Theory of Operation...
Page 58: ......
Page 62: ...Theory of Operation 3 4 WFM90D and WFM91D Service Manual...
Page 63: ...Performance Verification...
Page 64: ......
Page 82: ...Performance Verification Digital Section 4 18 WFM90D and WFM91D Service Manual...
Page 83: ...Adjustment Procedures...
Page 84: ......
Page 88: ...Adjustments 5 4 WFM90D and WFM91D Service Manual...
Page 98: ...Adjustment Procedures Digital Section 5 14 WFM90D and WFM91D Service Manual...
Page 99: ...Maintenance...
Page 100: ......
Page 130: ...Remove and Replace 6 30 WFM90D and WFM91D Service Manual...
Page 152: ...Troubleshooting Procedures 6 52 WFM90D and WFM91D Service Manual...
Page 153: ...Options...
Page 154: ......
Page 157: ...Replaceable Mechanical Parts...
Page 158: ......
Page 166: ...Replaceable Parts List 8 8 WFM90D and WFM91D Service Manual...
Page 167: ...Diagrams...
Page 168: ......
Page 170: ...Diagrams 9 2 WFM90D and WFM91D Service Manual...