Battery Pack Maintenance

Observe and note the run time that a new fully-charged

battery provides for powering your product. You can
use this new battery run time as a basis to compare run
times for older batteries. The run time of your battery
will vary depending on the product’s con

fi

guration and

the applications that you run.

The battery pack gives the best performance when operated

at normal room temperature, 20 °C ±5 °C (68 °F ±9 °F).

Routinely check the charge status of the battery.
Carefully monitor a battery that is approaching the end

of its estimated life.

Consider replacing the battery with a new one if the battery

run time drops below about 80% of the original run time,
or the battery charge time increases signi

fi

cantly.

Follow the storage requirements if you store or do not use

a battery for an extended period. If you do not follow the
storage requirements, and the battery will not power the
waveform monitor when installed, consider the battery
to be damaged. Do not connect power to the waveform
monitor during this time. Remove the battery pack from
the waveform monitor. Do not attempt to recharge or reuse
the battery. Replace it with a new battery.

Battery Pack Charging

The battery pack charges automatically when the waveform
monitor is connected to the external power supply. You can
also charge the battery pack with the Tektronix WFM200BC
External Battery Charger. Remove the battery pack from the
WFM200BC charger when charging is complete.
A Lithium-Ion battery pack will self-discharge during non-use.
To achieve the longest operating time, charge your battery
pack before use. If you plan to store battery packs, read
the instructions in the

Battery Pack Storage

section of this

document.
When the battery pack is in the WFM2200 waveform monitor,
the battery gauge (

) on the waveform monitor display

indicates the approximate amount of charge.

To avoid automatic shutdown of the battery pack, only run

the waveform monitor on the battery pack when the gauge
on the waveform monitor is green or yellow.

Always verify the correct placement of the battery in the

WFM2200 waveform monitor or the WFM200BC battery
charger.

When the battery pack is installed in the waveform

monitor, it will automatically charge whenever the
supplied AC adapter is connected, whether the instrument
is On, Off, or in Standby mode.

Con

fi

guration

Typical charging time

One fully discharged battery pack
charging in the WFM2200 waveform
monitor with the waveform monitor
turned on or off

5.5 hours

One or two fully discharged battery
packs charging in the WFM200BC
external battery charger

5.5 hours

CAUTION.

To avoid damage to the battery pack, use only

the WFM2200 waveform monitor or the WFM200BC
battery charger to charge the battery pack. Do not connect
any other voltage source to the battery pack.

NOTE.

For optimum performance, charge the battery pack

completely before using it for the

fi

rst time or after prolonged

storage.

To discharge the battery pack completely, continue to run
the waveform monitor on the battery pack until automatic
shutdown occurs. After the instrument shuts down, either
recharge the battery pack promptly or remove it from the
instrument until you are ready to recharge the battery pack.

NOTE.

To prolong the life of the battery pack and to prevent

shutdown, do not operate or charge the battery pack at high
temperatures. For best results, allow the battery pack to cool to
room temperature before using or charging the battery pack.

Do not leave a battery pack discharged for extended periods.
See the Storing Battery Packs section for information on how to
properly store a battery pack.

The battery pack stops charging under the following conditions:

The battery pack is fully charged.

The internal temperature of the battery pack exceeds the

safe charging threshold.

A fault condition occurs.

The battery pack shuts down and disconnects itself from the
WFM2200 waveform monitor under the following conditions:

The battery pack is fully discharged.

The internal temperature of the battery pack exceeds the

safe discharging threshold.

A fault condition occurs.

Battery Pack Handling

Do not disassemble, crush, drop, or puncture the battery.

Do not short the external contacts on the battery.

Do not dispose of the battery in

fi

re or water.

Do not expose the battery to temperatures above +60 °C

(+140 °F).

Seek medical advice if a battery or part of it has been

swallowed.

Do not put battery packs near heat or

fi

re. Do not put in

sunlight.

Keep the battery pack clean and dry. Clean dirty

connectors with a dry, clean cloth.

Batteries contain hazardous chemicals that can cause burns

or explode. If exposure to chemicals occurs, clean with
soap and water and get medical aid. Repair the product
before use if the battery leaks.

Do not attempt to open, modify, reform or repair a battery

pack that appears to be malfunctioning, or which has been
physically damaged.

Keep the battery away from children and animals.

Avoid exposing the battery to excessive shock or vibration.

Do not use a damaged battery or charger.

If a battery pack has leaking

fl

uids, do not touch any

fl

uids. Dispose of a leaking battery pack. See the

Battery Recycling

section for information on disposal and

recycling.

In case of eye contact with

fl

uid, do not rub eyes.

Immediately

fl

ush eyes thoroughly with water for at least

15 minutes, lifting upper and lower lids, until no evidence
of the

fl

uid remains. Seek medical attention.

Battery Pack Storage

Store battery packs in a low-humidity environment free of

corrosive gases. Storing battery packs in high-humidity
environments, or outside the temperature range, can
cause oxidation on the metallic parts, increased leakage,
signi

fi

cant aging and premature failure.

Do not store battery packs near heat or

fi

re. Do not store

in the sunlight.

Do not keep battery packs in a place where the terminals

can be shorted by metal objects (for example, coins, paper
clips, or pens).

Do not remove a battery pack from its original packaging

until required for use.

When possible remove the battery pack from the

equipment when not in use.

Charge the battery to approximately 50% of capacity at

least once every six months.

Remove the battery and store it separately from the

instrument. Refer to the

WFM2200 User Manual

for

instructions on removing and installing the battery.

For optimal battery life, store the battery at temperatures

b5 °C and +25 °C (+41 °F and +77 °F). Higher
temperatures reduce the battery storage life.

Battery Pack Replacement

Use the instructions to remove and replace battery packs.

NOTE.

Replace the Lithium-Ion battery pack only with a

WFM200BA battery pack.
See the Battery Recycling section for information on how to
properly dispose of a Lithium-Ion battery pack.

Speci

fi

cations

Characteristic

Description

Capacity, typical

6.14 Ah

Output, nominal

14.4 VDC

Charging temperature

Installed in the WFM2200 waveform
monitor or WFM200BC battery charger:
10 °C to +40 °C (+50 °F to +104 °F);
varies per the WFM2200 heat
dissipation characteristics, the actual
limit may be lower

Operating temperature

Installed in the WFM2200 waveform
monitor: 0 °C to +40 °C (+32 °F to
+104 °F)

Storage temperature

-20 °C to +60 °C (-4 °F to +140 °F)

Weight

Battery: 453.6 g (16.0 oz)
Battery (including packaging): 567 g
(20.0 oz)

Warranty Information

For warranty information, go to the www.tektronix.com Web
site, and search for “WFM200BA warranty”.

Contacting Tektronix

Tektronix, Inc.
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:

In North America, call 1-800-833-9200.
Worldwide, visit www.tektronix.com to

fi

nd contacts in

your area.

Copyright © Tektronix, Inc. All rights reserved.

www.tektronix.com

Reviews: