SCPI Commands
3–10
VX4237 Digital Multimeter Module User Manual
CALibration Subsystem
This subsystem is used to calibrate the ranges and functions of the DMM. This
will correct for any system errors due to drift or ageing effects.
Before any calibration can take place, two security levels must be set. First, there
is a switch on the DMM itself that must be set to CAL ENABLE. Having done
this, the command CALibration:SECure ON must be sent.
CALibration
:HIGH?
[<numeric_value>]
:LOW?
[<numeric_value>]
:SECure
<Boolean>
:SLFRequency?
There are no directly related commands however commands to configure the
DMM such as CONFigure, SENSe etc. are used in conjunction with CALibra-
tion.
See also Routine Calibration Procedure; section 8 of this handbook.
Syntax
Related Commands
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for VX4237
Page 5: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 29: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 39: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 145: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 161: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 187: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 200: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 207: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 208: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...