Maintenance
Maintenance
This section contains maintenance information for your probe.
Error condition
The TPR1000 and TPR4000 power rail probes are designed to work with all TekVPI-interface oscilloscopes and adapters.
However, there may be some cases where all of the probe features may not work properly.
If the Status LED is red or
fl
ashing during or after probe power on, an internal probe diagnostic fault exists. Disconnect
and reconnect the probe to restart the power-on diagnostic sequence. If the Status LED continues to be red or
fl
ashing
for more than 30 seconds when the oscilloscope application is running, the probe is defective and must be returned to
Tektronix for repair.
Replacement parts
There are no user replaceable parts within the probe. Refer to
Accessories
for a list of replaceable accessories for your probe.
Cleaning
Protect the probe from adverse weather conditions. The probe is not waterproof.
CAUTION.
To prevent damage to the probe, do not expose it to sprays, liquids, or solvents. Avoid getting moisture inside
the probe during exterior cleaning.
Do not use chemical cleaning agents; they may damage the probe. Avoid using chemicals that contain benzine, benzene,
toluene, xylene, acetone, or similar solvents.
Clean the exterior surfaces of the probe with a dry, lint-free cloth or a soft-bristle brush. If dirt remains, use a soft cloth or
swab dampened with a 75% isopropyl alcohol solution. A swab is useful for cleaning narrow spaces on the probe, use only
enough solution to dampen the swab or cloth. Do not use abrasive compounds on any part of the probe.
TPR1000 and TPR4000 User Manual