background image

Clear and Sanitize Procedures

Table 2: Nonvolatile memory devices (cont.)

Type and
minimum size

Function

User
modi

fi

able

Data
input
method

Location

Process to clear

To sanitize

PROM
1 M

Contains FPGA
image for
instrument control

No

Programmed
at the
factory, no
user data

Interface board

N/A. No user data stored
in this device.

N/A. No user
data stored in this
device.

PROM
4 M

Contains FPGA
image for
instrument control

No

Programmed
at the
factory, no
user data

Interface board

N/A. No user data stored
in this device.

N/A. No user
data stored in this
device.

Data Export Devices

The following table lists the data export devices in the standard instrument and
listed options.

Table 3: Data export devices

Type

Function

User
modi

fi

able

1

Input
method

Location

Process to disable

Analog outputs

Provides an analog copy of
signals from the device under
test

No

From the
device under
test

BNC connectors
on the side of the
instrument

N/A. Outputs cannot be disabled.

USB host port

Supports removable USB

fl

ash drive. User storage of

instrument setups and exported
waveform data.

Yes

Save

USB host ports on
front and rear of
instrument

Files can be deleted or overwritten on
the instrument or a PC; USB

fl

ash drive

can be removed and destroyed.
The USB ports can be disabled. (See
page 3,

Disabling USB and CD-RW

Capability

.)

CD-RW,
DVD-ROM
drive

Store and transport data

Yes

Read and
write

Drive on the side of
the instrument

Drive can be disabled. (See page 3,

Disabling USB and CD-RW Capability

.)

LAN Ethernet

Transfer data

Yes

Read, write,
and controls
programs

Ethernet port on
rear of instrument

Ethernet port can be disabled. (See
page 4,

To Disable LAN Ethernet

Connectivity Using the BIOS

.)

TekLink

Remote control of
instrumentation systems

Yes

Read, write,
and controls
programs

TekLink port on rear
of instrument

TekLink can be disabled. (See page 5,

To Disable TekLink Using the Windows
Device Manager

.)

1

During normal instrument operation.

Disabling USB and CD-RW

Capability

The following instructions give a method to disable the built-in USB and CD-RW
capability. Using the Windows Device Manager disables the devices for Windows
programs. This procedure disables both USB and the CD-RW to prevent their use.

NOTE.

If you disable the USB, CD-RW, and LAN in the following procedures,

you cannot write new

fi

rmware to the hard drive. To do so, you must enable one

of these items.

TLA6000 Declassi

fi

cation and Security Instructions

3

Summary of Contents for TLA6000 Series

Page 1: ...TLA6000 Series Logic Analyzers Declassification and Security ZZZ Instructions xx www tektronix com P077048300 077 0483 00...

Page 2: ...tion in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tektronix Inc TekLink i...

Page 3: ...Memory Devices 1 Data Export Devices 3 Built In Security Features 6 Troubleshooting 7 How to Clear or Sanitize a Nonfunctional Instrument 7 How to Recover from Clearing or Removing the Instrument s M...

Page 4: ...Table of Contents ii TLA6000 Declassification and Security Instructions...

Page 5: ...l TLA7ACx Logic Analyzer Module Service Manual Terms The following terms may be used in this document Clear This removes data on media memory before reusing it in a secured area All reusable memory is...

Page 6: ...temporarily or permanently from a secured area to a non secured area Scrub This is equivalent to sanitize User modifiable The memory device can be written to by the user during normal instrument opera...

Page 7: ...storing acquired data No Written by controller ASICs Acquisition board Remove power source from the instrument for at least 20 seconds Remove power source from the instrument for at least 20 seconds S...

Page 8: ...therboard Load new BIOS from vendor Web site Securely store instrument in a safe location or destroy it CMOS Configurable BIOS parameters No Written by the processor system Motherboard Remove battery...

Page 9: ...d rear of instrument Files can be deleted or overwritten on the instrument or a PC USB flash drive can be removed and destroyed The USB ports can be disabled See page 3 Disabling USB and CD RW Capabil...

Page 10: ...e Device Usage drop down list NOTE It is critical to leave the USB Root Hub operating otherwise the front panel will not function If the front panel does not function the TLA application will not run...

Page 11: ...ck the General tab and select Do not use this device disable in the Device Usage drop down list Removable Hard Disk Drive To Remove the Hard Disk Drive To remove the hard disk drive assembly refer to...

Page 12: ...f the hard disk drive has been removed and a new hard disk drive is installed you need to install the operating system and the TLA application software Refer to the TLA6000 Series Logic Analyzer Insta...

Page 13: ...tached USB flash drive from the instrument before returning the instrument for repair If your instrument is not functioning and you need to clear it remove the power source from the instrument for at...

Reviews: