Clear and Sanitize Procedures

Memory Devices

The following tables list the volatile and nonvolatile memory devices in the
standard instrument and options.

Table 1: Volatile memory devices

Type and
minimum size

Function

User
modi

fi

able

Data input
method

Location

To clear

To sanitize

DRAM,
1 M X 16

Firmware
execution code

No

Written by the
processor system

LPU board

Remove power
source from the
instrument for at
least 20 seconds.

Remove power
source from
the instrument
for at least
20 seconds.

DRAM,
1 M X 16

Acquisition data
transfer

No

Written by the
processor system

LPU board

Remove power
source from the
instrument for at
least 20 seconds.

Remove power
source from
the instrument
for at least
20 seconds.

BICMOS

Firmware
execution code

No

Written by the
processor system

LPU board

Remove power
source from the
instrument for at
least 20 seconds.

Remove power
source from
the instrument
for at least
20 seconds.

SDRAM
4 M X 16

Acquisition
memory for storing
acquired data

No

Written by
controller ASICs

Acquisition board

Remove power
source from the
instrument for at
least 20 seconds.

Remove power
source from
the instrument
for at least
20 seconds.

SDRAM
4 M X 16

Memory for
instrument code

No

Written by
controller ASICs

Interface board

Remove power
source from the
instrument for at
least 20 seconds.

Remove power
source from
the instrument
for at least
20 seconds.

SDRAM
1 M X 16

Acquisition
memory for storing
acquired data

No

Written by
controller ASICs

Acquisition board

Remove power
source from the
instrument for at
least 20 seconds.

Remove power
source from
the instrument
for at least
20 seconds.

SDRAM
16 M X 16

Acquisition
memory for storing
acquired data

No

Written by
controller ASICs

Acquisition board

Remove power
source from the
instrument for at
least 20 seconds.

Remove power
source from
the instrument
for at least
20 seconds.

ASIC
16 K

Processor
execution memory

No

Written by ASICs

Acquisition board

Remove power
source from the
instrument for at
least 20 seconds.

Remove power
source from
the instrument
for at least
20 seconds.

TLA6000 Declassi

fi

cation and Security Instructions

1

Summary of Contents for TLA6000 Series

Page 1: ...TLA6000 Series Logic Analyzers Declassification and Security ZZZ Instructions xx www tektronix com P077048300 077 0483 00...

Page 2: ...tion in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tektronix Inc TekLink i...

Page 3: ...Memory Devices 1 Data Export Devices 3 Built In Security Features 6 Troubleshooting 7 How to Clear or Sanitize a Nonfunctional Instrument 7 How to Recover from Clearing or Removing the Instrument s M...

Page 4: ...Table of Contents ii TLA6000 Declassification and Security Instructions...

Page 5: ...l TLA7ACx Logic Analyzer Module Service Manual Terms The following terms may be used in this document Clear This removes data on media memory before reusing it in a secured area All reusable memory is...

Page 6: ...temporarily or permanently from a secured area to a non secured area Scrub This is equivalent to sanitize User modifiable The memory device can be written to by the user during normal instrument opera...

Page 7: ...storing acquired data No Written by controller ASICs Acquisition board Remove power source from the instrument for at least 20 seconds Remove power source from the instrument for at least 20 seconds S...

Page 8: ...therboard Load new BIOS from vendor Web site Securely store instrument in a safe location or destroy it CMOS Configurable BIOS parameters No Written by the processor system Motherboard Remove battery...

Page 9: ...d rear of instrument Files can be deleted or overwritten on the instrument or a PC USB flash drive can be removed and destroyed The USB ports can be disabled See page 3 Disabling USB and CD RW Capabil...

Page 10: ...e Device Usage drop down list NOTE It is critical to leave the USB Root Hub operating otherwise the front panel will not function If the front panel does not function the TLA application will not run...

Page 11: ...ck the General tab and select Do not use this device disable in the Device Usage drop down list Removable Hard Disk Drive To Remove the Hard Disk Drive To remove the hard disk drive assembly refer to...

Page 12: ...f the hard disk drive has been removed and a new hard disk drive is installed you need to install the operating system and the TLA application software Refer to the TLA6000 Series Logic Analyzer Insta...

Page 13: ...tached USB flash drive from the instrument before returning the instrument for repair If your instrument is not functioning and you need to clear it remove the power source from the instrument for at...

Reviews: