Trigger Functions
Detect narrow pulses.
To set the instrument to trigger on narrow positive pulses
shorter than 5 ms, do the following:
1.
Press the
Trigger
button.
2.
Select
TRIGGER OPTIONS
.
3.
Select
Pulse Width on
and press the
Enter
button.
4.
Select the positive pulse icon ( ).
5.
Select the
=t (±10%)
condition.
6.
Select
On Trigger
for update. The instrument is now prepared to trigger on
narrow pulses only. A new menu will appear from which you can further
de
fi
ne settings.
In this screen, the pulse width was set to 4.80
μ
s. All positive pulses equal to
4.80
μ
s are displayed. (See Figure 36.)
Figure 36: Pulse width triggering
NOTE.
The instrument stores all triggered screens in the replay memory. For
example, if you set up your triggering for glitches, you can capture 100 glitches
with time stamps. Read more in the Replay section. (See page 52, Use Replay.)
70
THS3000 Series Oscilloscopes User Manual
Summary of Contents for THS3014
Page 1: ...xx THS3000 Series Oscilloscopes ZZZ User Manual P077060702 077 0607 02...
Page 2: ......
Page 3: ...THS3000 Series Oscilloscopes ZZZ User Manual xx www tektronix com 077 0607 02...
Page 6: ......
Page 14: ...General safety summary viii THS3000 Series Oscilloscopes User Manual...
Page 32: ...Introduction 16 THS3000 Series Oscilloscopes User Manual...
Page 96: ...Troubleshooting 80 THS3000 Series Oscilloscopes User Manual...
Page 112: ...Appendix A Specifications 96 THS3000 Series Oscilloscopes User Manual...
Page 118: ...Appendix B Connect the Instrument to a Computer 102 THS3000 Series Oscilloscopes User Manual...
Page 126: ...Appendix D Battery Pack 110 THS3000 Series Oscilloscopes User Manual...
Page 128: ...Appendix E Maintenance and Cleaning 112 THS3000 Series Oscilloscopes User Manual...