Adjustments
Table 12: Input LF-HF gain calibration points (cont.)
Cal step
UUT input value (5500A SCOPE, MODE volt, SCOPE Z 1 M
Ω
, 1 kHz)
CL 0601
50 mV
CL 0602
100 mV
CL 0603
250 mV
CL 0662
2 V
CL 0605
1 V
CL 0606
2.5 V
CL 0607
5 V
CL 0664
20 V
CL 0608
10 V
CL 0609
25 V
CL 0610
50 V (set 5500A to OPR!)
8.
Set the 5500A to Standby.
Input Position
Proceed as follows to do the Input Position calibration:
1.
Press F2 to select calibration adjustment step CL 0619.
2.
Remove all connections to the channels (open inputs).
3.
Press F3 to start the calibration.
4.
Wait until the display shows calibration status :READY.
5.
Proceed to the next step.
30
Technical Reference