![Tektronix ths3000 series Technical Reference Download Page 35](http://html1.mh-extra.com/html/tektronix/ths3000-series/ths3000-series_technical-reference_1079180035.webp)
Performance Veri
fi
cation
Test
Minimum
Incoming
Outgoing
Maximum
Input Phase Measurement
Accuracy
20 ms/div
–2 °
+2 °
200 ns/div
–2 °
+2 °
20 ns/div
–3 °
+3 °
Time Base
8.00 ms
-3 pixels
+3 pixels
800.0
μ
s
-3 pixels
+3 pixels
Trigger Sensitivity
100 mVpp @ 5 MHz
Stable trigger
—
400 mVpp @ 100 MHz
Stable trigger
—
800 mVpp @ 150 MHz
Stable trigger
—
400 mVpp @ 200 MHz
Stable trigger
—
800 mVpp @ 250 MHz
Stable trigger
—
AC Voltage Accuracy and
Bandwidth
1 MHz
835 mV
965 mV
25 MHz
790 mV
1.010 V
100 MHz
>630 mV
—
200 MHz
>630 mV
—
Video Trigger
Line 622 (PAL)/525 (NTSC)
Triggers on line
—
Line 310 (PAL)/262 (NTSC)
Triggers on line
—
Line 310 (PAL)/262 (NTSC)
<Neg Trig>
Triggers on line
—
Probe Calibration Generator
Pass/Fail
—
Technical Reference
23