THM500 Series Specifications
4
Bench Test Instruments and Handheld Oscilloscopes Basic Service
Table 2: DMM specifications (cont.)
565
560
550
General
Additional Readouts
MIN:
Minimum voltage or resistance
MAX:
Maximum voltage or resistance
MAX–MIN: Difference between MAX and MIN
HOLD:
Value of the main reading when the HOLD button is pressed
D
HOLD:
Difference between HOLD reading and active reading
D
D
D
DC voltage
Range and Resolution
Range
Resolution
400 mV
0.1 mV
4 V
1 mV
40 V
10 mV
400 V
100 mV
850 V
1 V
Autorange available; selects from all ranges except 400 mV.
D
D
D
Accuracy
±
(0.5% of r 5 counts)
D
D
D
Normal Mode Rejection
>60 dB typical at user selectable 50 or 60 Hz
D
D
D
Common Mode Rejection
>100 dB typical at user selectable 50 or 60 Hz
D
D
D
AC voltage
Range and Resolution (True RMS)
Range
Resolution
400 mV
0.1 mV
4 V
1 mV
40 V
10 mV
400 V
100 mV
600 V
1 V
Autorange not available on 400.0 mV range
D
D
D
Accuracy
±
(2% of r 5 counts) for 50 or 60 Hz sine wave. Add 2% of reading plus
5 counts for nonsinusoidal signal with crest factor <3.
D
D
D
W
/resistance
Range and Resolution
Range
Resolution
400
W
0.1
W
4 k
W
1
W
40 k
W
10
W
400 k
W
100
W
4 M
W
1 k
W
40 M
W
10 k
W
Autorange available
D
D
D
Accuracy
±
(0.5% of r 2 counts). 40 M
W
range is
±
(2% of r5 counts) at
≤
60% relative humidity.
Derate 4 M
W
range to
±
(1% of r 2 counts) from 60% to 90% relative
humidity.
D
D