Summary of Contents for TDS 340

Page 1: ...ss underutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing Rentals Demos...

Page 2: ...User Manual TDS 340 TDS 360 TDS 380 Digital Real Time Oscilloscopes 070 9459 01 1lllllI nlllllllllllllll Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 3: ...o any GTDEP user that is a full participant in the Metrology Data Interchange Data Base of GIDEP provided that all copies of the original work include the entire copyright notice and ownership stateme...

Page 4: ...f EdgeTriggering VidcoTriggering TaklngMeasurements f f Automated Measurements TatingMeasurements withCursors Controlling Acquisition _f Controlling the Display UsingthePFT _ Description Operation Con...

Page 5: ...ix B Performance Verification Conventions TeslEquipment TcstRccord PcrfOrmnceTestOvcrvicw SignalAcquisitionSystcrn Checks TimeBase SystemCheeks ig erSystem Checks SineWave Generator LcvclingProcedure...

Page 6: ...utton Figure 3 5 The horizontal menu Figure 34 The MATH button Figure 3 7 A math waveform Figure 3 8 The TRIGGER MENU button Figure 3 Y The edgetrigger menu Figure 3 10 The videotrigger menu Figure 3...

Page 7: ...g menu Figure 3 37 The UTILITY button Figure 3 38 The utility pop up menn Figure 3 3Y Dateand time display 3 35 3 36 3 38 3 39 341 342 342 3 15 345 3 48 3 H 3 51 3 53 3 57 3 59 3 60 361 Figure h l TDS...

Page 8: ...ng A 7 Table A 10 Nominal traits signal acquisition system A 7 Table A XX Nominal traits time basesystem A 8 Table A 12 Nominal traits triggering system A S Table A 13 Nominal traits display system A...

Page 9: ...ttime basesettingof 2 5nsidiv IDS 380 200 MHz analogbandwidthandfastesttime basesettingof 2 5 ns div IDS 360 100MHz analogbandwidthandfastesttime base settingof5 ns div TDS 340 A full complementof adv...

Page 10: ...ollows Smmx 20mm time lag 3 15A 250V or 1 25in x 0 25in cimc lag 3A 250V Connect theproper powercordfromtherear panel powerconnector Figurel 2 tothepowersystem Usethepowercordclamp tosecure the cordto...

Page 11: ...esoits bottom is down on thework SUrface andits rearis facing you 3 Find thelint cord on thercx cover SeeFigure l 2 Unplugtheline cord from its reccptaclc 4 Find thefusedrawerbeneaththeline voltageplu...

Page 12: ...1 4 TDs 340 ms 360 IDS 380 user Manual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 13: ...heside menubutton OK Confirm RunTest The internaldiagnosticsverify properoscilloscope function This vcrlfication takesabout30 seconds While it progresses a variety of testpatternsflashon screen Whenfi...

Page 14: ...peasate S II Paths When compensation completes thestatus message updates to Pass orFailinthemaan menu 9 CheckthatthewordPassappears underSignal Pathinthemainmt nu See Figure l 3 If Passdoesnotappear t...

Page 15: ...eclledEnthepeflormancetests which beginon pug eB 5 110NOT makechangesto thefront penel settingsthat arenot calledout in the procedure If youmakechangesto thesesettingsotherthanthoseculledout in thepro...

Page 16: ...wave signal asyourotateit Leave thesignal untriggcred 12 Checkthatpressing SETLEVELTO50 triggersthesignal thatyoujust left untriggcred 13 Press thefront panel buttonACQUIRE themain menu buttonMode and...

Page 17: ...TDS 360 TDS 380 21 Disconnect terminal UserManual theprobefrom thechannel input andthePROBE GettingStxtcd COMP Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg...

Page 18: ...ubuttaxiprovide TheSideMenubuttaxiprovide access tosidemenusalections access tosidemenusalections Seepage2 9 form0rf z Seepage2 9 form0rf z information aboutIho user information aboutIho user interfac...

Page 19: ...TheVER iiCAtMENUbunoncallsup theverticaloperadons menu Forrnoro information abati vertical operations seepagez 3 The SCALEknobcontrols thevefliwl scaleofthesolededwaveform Probecornpens output see pag...

Page 20: ...buttonM ISup iho horizontal operationsmsnu Formore information abouthorizontal operations see page SC TheSCALE knobrnntrolsthehorizontal scaleoftheactivewaveforms IDS 340 TDS 360 S380UscrManual 2 3 Ar...

Page 21: ...General PurQOSe Knob controls many side menu fundions including the cursors The SELECT button switches control from cursor to 0 rscI The SAWRECALL button calls up the save recall me Seepage345for more...

Page 22: ...here Cursormeasurement readouts SeepageZ 21 formoreinformation about C rsOrs Thesidemew offersa choiceofspecific actions TheTriggerreadoulshows thetriggerSDU W andlevel andwhetherlheuscilloscope istri...

Page 23: ...ignals for display TheEXTTRIGinputacceptsex emai triggersignals Seepage3 12 for moreinformation aboutexternal triggering lB 7 340 TDS 360 TDS 380 User Manual Artisan Technology Group Quality Instrumen...

Page 24: ...2 and GPIB interfaces to Iranemil hardcopy data see page Wl tar hard copy procedures You can use lhe WI6 and RS 232 innixes to operateaodprogt mtheoscilloscopefromaGPl controller see Ih TDS 340 lDS 36...

Page 25: ...t hxe The procedureon page2 10 illustrateshow to navigatein themenustructure If you areunfamiliar with this menusystem you may want to run through the procedureseveraltimesto learnhow you canaccessfun...

Page 26: ...ifdisplayed Figure2 1 Usingmenus 4 Adjud menu itam values with general purpose knob 1 Pushone of theindicatedfront panelbuttonsto call up a menuOSfunctions This first menuis themain menu Sometimesthe...

Page 27: ...tom menubuttonsometimesactivatesa pop upmenu asshownin Figure 2 2 You cancycle throughthepop upmenu optionsby repeatedlypressingthebutton Eachselectioncrllls up different bottom andsidemenus Figure 2...

Page 28: ...hepro bc to her theCH 1orCA2inpul Besuretotwisttheprobe endclockwise untiltie BNC isfully locked See Figure 2 3 Figure2 3 Connectinga probe Usethefollowingprocedure tocompensate theprobeeverytim cyous...

Page 29: ...thecompensation waveform It shouldresemblethewaveformshown in Figure2 5 I I 1 II II II II II 0 0 0110 0110 0 Figure2 5 Compensation waveform TDS 340 TDS 360 TDS 380UserManual 2 I 3 Artisan Technology...

Page 30: ...tiigkcnt tool ProvidCdwith theprobe10adjustthecompensationasshownin Figure z 7 1 L Probecompensated correctly n Probeundorcompensaied ns _ Figure 2 6 How Probe compensation affects signals Figure Z 7...

Page 31: ...played channels Acauire Mode I Sam Acquire StopAfter DisplayStyle Disolav lntensitv Overall RUN STOPbutton only vectors If less than 50 set to75 Display Format Horizontal Position Yr Centered within t...

Page 32: ...oldoff VerkcalScale LYF _I VerticalCoupling vet iCalBandwidth VerticalOffset Asdehmined bythe signallevel DCunlessAC waspreviouslyset AC remainsunchanged Full 0 volts 2 16 Artisan Technology Group Qua...

Page 33: ...eraturc between 20 C and 30 C hashad a warn up periodof atleast20 minntcs andis operatingatan ambienttemperaturebctwcen 10 C and 55 C unless otherwisenoted 0 I C 7 Table A l Warranted characteristics...

Page 34: ...ge level at the center of the A D converter dynamic range Offset Accuracy is the accuracy of this voltage level 2 The AC Coupled Lower Frequency limits are reduced by a factor of IOwhen IOX passive pr...

Page 35: ...increasingto5OOmVat4OOMHz TDS 380 50 mV from DC to 50 MHz xtemal l 0 TDS 340 500 mV from DC to 20 MHt increasingto 1 5Vat 100 MHz TDS 360 500 mV from DCto 50 MHz increasingto 1 5 Vat 200 IMHz TDS 380...

Page 36: ...oppy disk drive t10 C to GO0 C operating or non operating lelative humiditywithoutdiskette in floppy diskdrive to80 at or below 29 C or to 20 fmm 30 C tot50 C operating to90 at or below t40 C or to 5...

Page 37: ...20 MHz Eand width Limited 20 MHz StepResponse SettlingError VoiWDiv step Settling error 3 setting amplitude 100 ns 20ms 2 mV div 99 5 mV div 2V 51 o 50 1 100mV div 995 mV div 520 V 21 5 so 2 1Vldiv I...

Page 38: ...int must not switch between opposite slopes on the waveform and the display must not roll across the screen on successive acquisitions The TRIG D LEDstays constantly lighted when the SEC D DIV setting...

Page 39: ...ofluoride battery Nominal traits aredescribedusingsimple statementsof fact suchas Two identical for thetrait Input Channels Numberof ratherthanin termsof limits thatareperformdncc requirements Nominal...

Page 40: ...the real time sample rate for awaveform created by interpolation of real time samples from a single acquisition or by equivalent time sampling of multiple acquisitions the waveform rate is faster tha...

Page 41: ...ieswith 1IEEEStd488 1987 I Partof Option 14 110interfaceorTD3F14A l O interfacefield upgradekit a g pinmale DTERS 232 interfacethat complieswith EWTIA 574 90 Partof Option 14 110intetface or T03F14A l...

Page 42: ...Dimensions StandardInstmment Figure A l 4 5 kg 10 Ibs 7 6 kg 17 5 Ibs when kitis packaged fordomestic shipmen Height 191 mm 7 5in with feet and accessoriespouchinstalled 165 mm 6 5in without the acces...

Page 43: ...ndix A Specilications Figure A i TDS 340 TDS 360 and TDS 380 dimensions TDS 340 TDS 360 TDS 380 User Manual A 11 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg...

Page 44: ...sionwaveform displacement orSO 6divisionincreaseinp p noisefrom 27 MHzto 500 MHz Testconditions both channelinputsterminatedwith grounding caps both channelsset to 10mV div both channelsset to DCCoupl...

Page 45: ...oadditionalequipmentotherthana StandaId aCCCSSOry probe and morecompletelytesttheinternalhardwareof this oscilloscope Youcanuse them to quickly determineif theoscilloscopeis suitablefor pulting into s...

Page 46: ...ifyareadout orstatus message thename of thebuttonor knobappears inboldface type m Inst ctionsfor menu selection followthisformat FRONTIPANEL EU1TON Pop Up if necessary Main MenuButton SideMenu Button...

Page 47: ...ccuracy 7 Generator 200 kHzto 250 MHz variable Wavetek9100 Calibration Checking bandwidthand trigger LeveledSineWave amplitudefrom 5 mV to4 Vpp Systemwith Option 250 sensitivity Medium Frequency into5...

Page 48: ...sample rate and delay time accuracy 2 0 Div 2 0 Div Edgetrigger sensitivity DC coupled MainTtigger stabletrigger N A MainTrigger Falling stable trigger N A 1 Generator set at 0 6 V 2 Generator set at...

Page 49: ...5S C Signal Acquisition System Checks Theseprocedures checksignalacquisitionsystemcharacteristics thatarelistedas checkedunderWurrantedCharacteris6ics in theSpecificationssccl ion CheckDCVoltage Measu...

Page 50: ...nu labelMeanappears inthe sidemenu Press thesidemenubuttonMean 7 Setthevertical SCALE to oneof thesettings listedinTable E 2thatyou havenot yet checked Start with the first setting listed 8 Press VERT...

Page 51: ...vertical scaleandposition offsct gcn eratorsettings 1 4 Repeatsteps7 through 13until youhavecheckedall thevertical scale settingslisted in TableB 2 Recordthemeasurements for eachof the50 mV settings 1...

Page 52: ...ion on obtaininga leveledoutput m an unlcvekd sineuwe generam Leveled Sine Wave Generator L owul b SOR Terminalion FigureB 3 Hookupfor analogbandwidthcheck 1 Press SAVE lZECAI L SETUP RecallFactorySet...

Page 53: ...ton screen asshownin Figure B l is 2 42 5mV First increase the reference frequency IDrhe fesl frequency then decrease the horizontal scale Second read Ihe resutls from the readout of measuremenf Pk Pk...

Page 54: ...s Prerequisites See page B 5 Procedure 1 Connect through a50Q precision coaxial cable anda50Q termination the time mtukoutputof atime marker generator toCH 1 asshown inFibwe B 5 Settheoutputof thegene...

Page 55: ...line at apoint within ti 0 divisions of thegratlculecerner NOTE Onedivision ofdisplacementfrom thecentergraticule correspondsto a SOppm time baseerrox 12 Disconnectthetesthookup DeltaTimeMeasurement A...

Page 56: ...ure B 6 ToEfl Trigger 50 n Terrninah Figure B 6 Hookup for trigger sensitivity check 8 Scrthegenerator frequency to 100MHZ Trx 340 200MH TDS360 or 400 MHZ TDS 3SO 9 Press MEASURE t High LowSetup Min M...

Page 57: ...sTRIGGER MENU Source Ch2 19 Disconnectthehookupfrom CEI 1 andcormectit to CH 2 20 SettheverticalSCALE to 500 mV ciiv 21 RepeatstepsI4 and 15for Channel2 22 PressTRIGGER MENU 3 Source EXT lO 23 PressME...

Page 58: ...e Note thereadingon thelevel meter Changethesinewavegeneratorto thedesirednewfrequency Input thecorrectionfactorfor thenewfrequencyinto thelevel meter djuS the sine wave _ cncrabr amplitude until the...

Page 59: ...dix 6 Pcrfomance Vetification FigureB 8 Hookupfarsinewavegenerator leveling IDS 340 TDS 360 8 TDS 380 UserManual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg...

Page 60: ...sedisplaymanufacturersusediffcrcnt pin combinationsandconnectors you may find Iheinformation in TableC l helpful Table C l VGA output connector pins Pin Signal 2 1Video monochrome analog 4 I 1Horizont...

Page 61: ...O I33 xy 070 9440 Xx 070 9441 XX 070 Q 7 xx LR LQ Standarc Chinese Korean 070 9438 Xx 070 9439xX Thek hwin options addtotheservices available withthesYdndard warranty Thestandard warranty appears imme...

Page 62: ...Part number C9 Option4 includesAdapter Hood 016 1154 01 OscilloscopeCart Rackmount Kit for field conversion 016 1166 00 Soft Sided CarryingCase 016 1156 01 CarryingCase 016 0792 01 Deluxe TransitCase...

Page 63: ...ssorycables Cable type GPIE 1 meter 3 3 feet GPIE 2 meter 6 6feet K 232 e pin female to g pin female connectors null modem 76 inch for ATstylecomputers RS 232 g pin female to25pin female connectors nu...

Page 64: ...hefollowing steps 1 2 Removelooseduston theoutsideof theoscilloscopewith a lint freecloth Usecareto avoidscratchingtic clearplasticdisplayfilter Usea soft cloth or papertowel dampcncdwith waterto clea...

Page 65: ...uipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentat...

Reviews: