Ia
Performance
Verification
R
R
test Equipment
These procedures use external, traceable signal sources to directly check
instrument
performance. If your test equipment does not meet the minimum
requirements listed in Table 6-1, your test results will be invalid.
Table 6-1: Test Equipment
Item Number and
Minimum Requirements
Example
Purpose
Description
1
Termination
Impedance 500; connectors:
Tektronix part number
Checking delay
50 0
female BNC input, male BNC
011-0049-01
between channels
(two required)
output
2
Cable, Precision
500, 91 cm (36 in), male to
Tektronix part number
Signal
Coaxial (two re-
male BNC connectors
012-0482-00
interconnection
quired)
3
Connector,
Female-BNC to dual-banana
Tektronix part number
Several accuracy
Dual-Banana
103-009-00
tests
4
Connector,
Male-BNC to dual-female-
Tektronix part number
Checking trigger
BNC “T”
BNC
103-0030-00
sensitivity
5
Coupler,
Female-BNC to dual-male-
Tektronix part number
Checking delay
Dual-Input
BNC
067-0525-02
between channels
6
Generator, DC
Variable amplitude to
±
110 V;
Data Precision 8200
Checking DC offset,
Calibration
accuracy to 0.1%
gain, and
measurement
accuracy
7
Generator,
200 khz to 250 MHz; variable
TEKTRONIX SO 503 Leveled
Checking trigger
Leveled Sine
Wave, Medium-
amplitude from 5 my to 4 V2.~, Sine Wave Generator
sensitivity at low
into 50 0
•
frequencies
Frequency
8
Generator, Time
Variable marker frequency
TEKTRONIX TO SO1A Time
Checking sample
Mark
from 10 ms to iOns;
Mark Generator
rate and delay-time
accuracy within 2 ppm
accuracy
9
Probe, lOX,
A P6109B (TDS 310 and
Tektronix number
Signal interconnec-
included with
TDS 320) or P6111 B
P61098 (TDS 310 and
tion
this instrument
(TDS 350) probe
TDS 320) or
P6111B(TDS 350)
Test Record
Photocopy the next page and use it to record the performance test results
for your instrument.
itS 310, TDS 320, & TEDS 350 Instruction Manual