Environmental Characteristics
The environmental characteristics are warranted performance speci
fi
cations. (See
Unlike the warranted characteristics, the environmental characteristics are type
tested; therefore there are no performance veri
fi
cation procedures provided to test
these characteristics. Unless otherwise noted, these characteristics apply to all
probes and ampli
fi
ers. (See page 49,
Table 13: Environmental characteristics
Parameter
Characteristic
Ambient Temperature
Operating
0 °C to +50 °C (32 °F to 122 °F)
Nonoperating
-40 °C to +75 °C (-40 °F to 167 °F)
Humidity
Operating
5 to 95% R.H. to +30 °C (86 °F)
5 to 85% R.H. +30 °C to +50 °C (86 °F to 122 °F)
Nonoperating
5 to 95% R.H. to +30 °C (86 °F)
5 to 85% R.H. +30 °C to +75 °C (86 °F to 167 °F)
Altitude
Operating
3000 m (9842 ft) maximum
Nonoperating
12,192 m (40,000 ft) maximum
Random Vibration, Ampli
fi
ers
Operating
0.31 g
RMS
, 5 Hz to 500 Hz, 10 minutes each axis
Nonoperating
2.46 g
RMS
, 5 Hz to 500 Hz, 10 minutes each axis
Random Vibration, Probes
Operating
2.66 g
RMS
, 5 Hz to 500 Hz, 10 minutes each axis
Nonoperating
3.48 g
RMS
, 5 Hz to 500 Hz, 10 minutes each axis
Shock, Ampli
fi
ers
50 g, 11 ms duration, half-sine pulses
Electro-Magnetic Compliance
Meets FCC Part 15, Subpart B, Class A
TCPA300/400 Ampli
fi
ers and TCP300A/400 Series Current Probes User Manual
53