Speci
fi
cations
These speci
fi
cations are valid only under the following conditions:
The probe and ampli
fi
er have been calibrated at an ambient temperature of
23 ° ±5 °C.
The probe and ampli
fi
er are operating within the environmental limits
described in Table 4-6 on page 4-5 (the operating temperature limits are 0 °C
to +50 °C, unless otherwise stated).
The probe and ampli
fi
er have had a warm-up period of at least 20 minutes.
The probe degauss/autobalance routine has been performed after the
20-minute warm-up period, and thereafter whenever the PROBE
DEGAUSS/AUTOBALANCE light blinks.
The ampli
fi
er output is properly terminated into 50
Ω
.
Speci
fi
cations are separated into two categories: warranted speci
fi
cations and
nominal or typical characteristics.
Warranted Speci
fi
cations
Warranted speci
fi
cations, Table 4-1, are guaranteed performance speci
fi
cations
unless speci
fi
cally designated as typical or nominal.
Table 8: Warranted TCPA300 and TCPA400 speci
fi
cations
Ampli
fi
er
TCPA300
TCPA400
DC Gain Accuracy
≤
1%
≤
1%
Installed probe
TCP312A
TCP305A
TCP303
TCP404XL
Bandwidth (-3 dB)
DC to 100 MHz
DC to 50 MHz
DC to 15 MHz
DC to 2 MHz
Rise Time, 10% to
90%
≤
3.5 ns
≤
7 ns
≤
23 ns
≤
175 ns
DC Gain Accuracy:
Warranted
≤
3%
≤
3%
≤
3%
1
≤
3%
Typical
2
≤
1%
≤
1%
≤
1%
≤
1%
1
Warranted from 10 °C to 50 °C. For temperature range of 0 °C to <10 °C, spec is +3%/-6%.
2
23 °C ±5 °C
TCPA300/400 Ampli
fi
ers and TCP300A/400 Series Current Probes User Manual
49