Troubleshooting and Error Codes
Table 6: Troubleshooting (cont.)
Problem
Remedy
Measurement aberrations exceed the
speci
fi
ed limit
The ampli
fi
er output is not terminated into 50
Ω
load. Set the input impedance of the
oscilloscope to 50
Ω
or connect a 50
Ω
feedthrough termination at the oscilloscope input.
Do not attach the termination to ampli
fi
er output.
The measurement exceeds the maximum continuous current or Ampere-second product
ratings of the Current Probe. (Current limits are available) (See page 27,
Excessive low frequency droop in pulse
response or low DC gain accuracy
Check that the probe slider is fully closed and locked.
Check and clean the probe transformer and lid.
Degauss the probe.
The ampli
fi
er or the oscilloscope may be AC coupled. Set the COUPLING to DC.
Measurements exhibit excessive delay
or slowed pulse response
The ampli
fi
er output is not terminated into 50
Ω
load. Set the input impedance of the
oscilloscope to 50
Ω
or connect a 50
Ω
feedthrough termination at the oscilloscope input.
Do not attach the termination to the ampli
fi
er output.
The measurement exceeds the Ampere-second product of the Current Probe.
The oscilloscope bandwidth limit is turned on. Verify that the bandwidth limit switch on
the oscilloscope is set to the desired bandwidth position.
44
TCPA300/400 Ampli
fi
ers and TCP300A/400 Series Current Probes User Manual