Terms used in this document

The following terms may be used in this document:

Clear.

 This removes data on media/memory before reusing it in a secured

area. All reusable memory is cleared to deny access to previously stored

information by standard means of access.

Erase.

 This is equivalent to clear.

Instrument Declassification.

 A term that refers to procedures that must be

undertaken before an instrument can be removed from a secure environment.

Declassification procedures include memory sanitization and memory

removal, and sometimes both.

Media storage/data export device.

 Any of several devices that can be used

to store or export data from the instrument, such as a USB port.

Nonvolatile memory.

 Data is retained when the instrument is powered off.

Power off.

 Some instruments have a “Standby” mode, in which power is still

supplied to the instrument. For the purpose of clearing data, putting the

instrument in Standby mode does not qualify as powering off. For these

products, you will need to either press a rear-panel OFF switch or remove the

power source from the instrument.

Remove.

 This is a physical means to clear the data by removing the memory

device from the instrument. Instructions are available in the product Service

Manual.

Sanitize.

 This eradicates the data from media/memory so that the data cannot

be recovered by other means or technology. This is typically used when the

device will be moved (temporarily or permanently) from a secured area to a

non-secured area.

Scrub.

 This is equivalent to sanitize.

User Accessible.

 User is able to directly retrieve the memory device

contents.

User-modifiable.

 User can write to the memory device during normal

instrument operation, using the instrument interface or remote control.

Volatile memory.

 Data is lost when the instrument is powered off.

 

Preface

Declassification and Security Instructions 

v

Summary of Contents for TBS1022

Page 1: ...TBS1000 TBS1000B TBS1000B EDU TDS1000B TDS2000B TDS1000C EDU TDS2000C and TPS2000B Series Digital Storage Oscilloscope Declassification and Security Instructions P077077102 077 0771 02...

Page 2: ......

Page 3: ...TBS1000 TBS1000B TBS1000B EDU TDS1000B TDS2000B TDS1000C EDU TDS2000C and TPS2000B Series Digital Storage Oscilloscope Declassification and Security Instructions www tektronix com 077 0771 02...

Page 4: ...and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tek...

Page 5: ...nvolatile memory devices container TBS1000B TBS1000B EDU 5 Data export devices 6 Clear and sanitize procedures Clear the flash memory 7 Clear or sanitize a non functional instrument Acquisition board...

Page 6: ...Table of Contents ii Declassification and Security Instructions...

Page 7: ...dures in this document are written to meet the requirements specified in NISPOM DoD 5220 22 M Chapter 8 ISFO Process Manual for Certification Accreditation of Classified Systems under NISPOM Products...

Page 8: ...number 077 0772 xx TBS1000B Series Digital Storage Oscilloscope Service Manual Tektronix part number 077 0897 xx TDS1000B and TDS2000B Series Digital Storage Oscilloscope Service Manual Tektronix part...

Page 9: ...upplied to the instrument For the purpose of clearing data putting the instrument in Standby mode does not qualify as powering off For these products you will need to either press a rear panel OFF swi...

Page 10: ...nged by the user Both Describes the type of information stored in the device It means that both user data and user settings are stored in the device None Describes the type of information stored in th...

Page 11: ...o sanitize Remove power from the instrument for at least 20 seconds SRAM 128 K Type and size SRAM 128 K X 8 Function Shared memory between the acquisition system and the USB processor Type of user inf...

Page 12: ...only User accessible No To clear Remove power from the instrument for at least 20 seconds Process to sanitize Remove power from the instrument for at least 20 seconds SRAM 64 K Type and size SRAM 64 K...

Page 13: ...ard User accessible No To clear Remove power from the instrument for at least 20 seconds Process to sanitize Remove power from the instrument for at least 20 seconds LPDDR SDRAM Type and size 1 g bit...

Page 14: ...pgrade Location Main Acquisition board User accessible No To clear Perform the procedure Clear the flash memory Process to sanitize Perform the procedure Clear the flash memory EEPROM Type and size EE...

Page 15: ...are operations Location Main Acquisition board User accessible No To clear Remove power from the instrument for at least 20 seconds Process to sanitize Remove power from the instrument for at least 20...

Page 16: ...sabled USB device port Type and size USB device port Function Supports remote control and data transfer to a PC Method of modification Directly Data input method Remote control via USBTMC Location Rea...

Page 17: ...Save To option button until Ref is selected Push the Source option button until CH1 is selected 11 Push the To option button until RefA is selected Push the Save option button to overwrite reference w...

Page 18: ...the hard copy image file with an image that does not contain any useful information 21 Power off the oscilloscope and then power on the oscilloscope to complete the process Clear and sanitize procedu...

Page 19: ...r company s internal policies regarding handling or disposal of the board Compact flash card TPS oscilloscopes only Remove the compact flash card and return the instrument to Tektronix for repair Afte...

Page 20: ...Clear or sanitize a non functional instrument 10 Declassification and Security Instructions...

Reviews:

Related manuals for TBS1022