Data export devices

USB host port

Type and size

USB host port

Function

Supports removable USB flash drive. User storage of reference waveforms, screen images, and
instrument setups.

Method of modification

Directly

Data input method

Save

Location

Front panel

To disable

Files can be deleted or overwritten on the oscilloscope or a PC; USB flash drive can be removed
and destroyed. The USB host port cannot be disabled.

USB device port

Type and size

USB device port

Function

Supports remote control and data transfer to a PC

Method of modification

Directly

Data input method

Remote control via USBTMC

Location

Rear panel

To disable

The USB device port cannot be disabled.

Compact flash card

Type and size

Compact flash card (TPS only)

Function

User storage of reference waveforms, screen images, and instrument setups

Method of modification

Directly

Data input method

By system resources

Location

Front panel

To disable

The Compact Flash port cannot be disabled.

Memory devices

 

6

 Declassification and Security Instructions

Summary of Contents for TBS1022

Page 1: ...TBS1000 TBS1000B TBS1000B EDU TDS1000B TDS2000B TDS1000C EDU TDS2000C and TPS2000B Series Digital Storage Oscilloscope Declassification and Security Instructions P077077102 077 0771 02...

Page 2: ......

Page 3: ...TBS1000 TBS1000B TBS1000B EDU TDS1000B TDS2000B TDS1000C EDU TDS2000C and TPS2000B Series Digital Storage Oscilloscope Declassification and Security Instructions www tektronix com 077 0771 02...

Page 4: ...and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tek...

Page 5: ...nvolatile memory devices container TBS1000B TBS1000B EDU 5 Data export devices 6 Clear and sanitize procedures Clear the flash memory 7 Clear or sanitize a non functional instrument Acquisition board...

Page 6: ...Table of Contents ii Declassification and Security Instructions...

Page 7: ...dures in this document are written to meet the requirements specified in NISPOM DoD 5220 22 M Chapter 8 ISFO Process Manual for Certification Accreditation of Classified Systems under NISPOM Products...

Page 8: ...number 077 0772 xx TBS1000B Series Digital Storage Oscilloscope Service Manual Tektronix part number 077 0897 xx TDS1000B and TDS2000B Series Digital Storage Oscilloscope Service Manual Tektronix part...

Page 9: ...upplied to the instrument For the purpose of clearing data putting the instrument in Standby mode does not qualify as powering off For these products you will need to either press a rear panel OFF swi...

Page 10: ...nged by the user Both Describes the type of information stored in the device It means that both user data and user settings are stored in the device None Describes the type of information stored in th...

Page 11: ...o sanitize Remove power from the instrument for at least 20 seconds SRAM 128 K Type and size SRAM 128 K X 8 Function Shared memory between the acquisition system and the USB processor Type of user inf...

Page 12: ...only User accessible No To clear Remove power from the instrument for at least 20 seconds Process to sanitize Remove power from the instrument for at least 20 seconds SRAM 64 K Type and size SRAM 64 K...

Page 13: ...ard User accessible No To clear Remove power from the instrument for at least 20 seconds Process to sanitize Remove power from the instrument for at least 20 seconds LPDDR SDRAM Type and size 1 g bit...

Page 14: ...pgrade Location Main Acquisition board User accessible No To clear Perform the procedure Clear the flash memory Process to sanitize Perform the procedure Clear the flash memory EEPROM Type and size EE...

Page 15: ...are operations Location Main Acquisition board User accessible No To clear Remove power from the instrument for at least 20 seconds Process to sanitize Remove power from the instrument for at least 20...

Page 16: ...sabled USB device port Type and size USB device port Function Supports remote control and data transfer to a PC Method of modification Directly Data input method Remote control via USBTMC Location Rea...

Page 17: ...Save To option button until Ref is selected Push the Source option button until CH1 is selected 11 Push the To option button until RefA is selected Push the Save option button to overwrite reference w...

Page 18: ...the hard copy image file with an image that does not contain any useful information 21 Power off the oscilloscope and then power on the oscilloscope to complete the process Clear and sanitize procedu...

Page 19: ...r company s internal policies regarding handling or disposal of the board Compact flash card TPS oscilloscopes only Remove the compact flash card and return the instrument to Tektronix for repair Afte...

Page 20: ...Clear or sanitize a non functional instrument 10 Declassification and Security Instructions...

Reviews:

Related manuals for TBS1022