Speci
fi
cations
Speci
fi
cations
All speci
fi
cations apply to the TBS1000 series models. TPP0101 and TPP0201
probe speci
fi
cations appear in Appendix B. To verify that the oscilloscope meets
speci
fi
cations, the oscilloscope must
fi
rst meet the following conditions:
The oscilloscope must have been operating continuously for twenty minutes
within the speci
fi
ed operating temperature.
You must perform the Do Self Cal operation, accessible through the Utility
menu, if the operating temperature changes by more than 5 °C (9 °F ).
The oscilloscope must be within the factory calibration interval.
All speci
fi
cations are guaranteed unless noted “typical.”
Oscilloscope Speci
fi
cations
Table 1: Input Speci
fi
cations
Characteristic
Description
Maximum Input Voltage
At front panel connector, 300 V RMS, Installation Category II; derate at 20 dB/decade above 100 kHz to 13 V peak AC at 3 MHz and
above.
Based upon sinusoidal or DC input signal. Maximum viewable signal while DC coupled is ±50 V offset ±5 V/division at 4 divisions, or
70 V. AC coupling allows measuring signals on a DC level up to 300 V. For non-sinusoidal waveforms, peak value must be less than
450 V. Excursion above 300 V should be less than 100 ms duration and the duty factor is limited to < 44%. RMS signal level must be
limited to 300 V. If these values are exceeded, damage to the instrument may result.
Table 2: General Speci
fi
cations
Characteristic
Description
Probe Compensator Output
Probe Compensator, Output
Voltage and Frequency, typical
Characteristics are as follows:
Output voltage 5.0 V ±10% into 1 Meg
Ω
load
Frequency 1 kHz
Power Source
Source Voltage
Full Range: 100 to 240 VAC RMS ±10%, Installation Category II (Covers range of 90 to 264 VAC); 50/60 Hz.
115 VAC RMS ±10%; 400 Hz.
Power Consumption
Less than 30 W at 90 to 264 VAC input.
Environmental
Operating
0° C to +50° C, with 5° C/minute
maximum gradient, non-condensing,
up to 3000m altitude
Temperature
Nonoperating
-40° C to +71° C, with 5° C/minute
maximum gradient
Cooling Method
Convection cooled
TBS1000 Series Oscilloscopes Installation and Safety Manual
27