![Tektronix RSA5100A Series Quick Start User Manual Download Page 51](http://html1.mh-extra.com/html/tektronix/rsa5100a-series/rsa5100a-series_quick-start-user-manual_1078155051.webp)
Operation
Connecting a Signal
The analyzer has a single RF signal input located on the front panel. Make sure that you observe the input signal
requirements before connecting a signal to the input.
CAUTION.
To prevent static damage to the instrument or to any instrument accessories, discharge to ground any
electrostatic charge that may be present on the center and outer conductors of cables, before attaching the cables to the
instrument inputs. Do not create an ESD (electrical static damage) antenna by leaving cables attached to the instrument with
the other ends of the cables open.
Required Signal Levels
The following table shows the input signal levels over which the analyzer can be used for measurements. The accuracy is
guaranteed at a signal amplitude and frequency (normalization point). Measurements can be made on signals within the
entire range of allowable inputs, but the accuracy of the measurement is affected by the frequency and amplitude of the
signal to be measured (see the RSA6100A Series or RSA5100A Series data sheet for more information).
RF signal input range (CW or peak envelope power)
Modulation measurement requirement
Spectral display requirement
-50 dBm to +30 dBm
DANL to +30 dBm
Input signal requirements
Characteristic
Description
Input impedance
50
Ω
Input frequency range
RSA6106A
RSA6114A
RSA6120A
RSA5103A
RSA5106A
RSA5103A/RSA5106A
9 kHz to 6.2 GHz
9 kHz to 14 GHz
9 kHz to 20 GHz
9 kHz to 3 GHz (RF Band)
9 kHz to 6.2 GHz (RF Band)
1 Hz to 32 MHz (LF Band)
Maximum measurable continuous input power
1 W (+30 dBm)
Maximum voltage rating
RSA6100A: ±40 V DC, AC coupled
RSA5100A – ±5 V DC, AC coupled
CAUTION.
To prevent damage to the instrument, do not connect CW signals greater than 1 Watt (+30 dBm). For the
RSA6100A Series, the maximum voltage rating is ±40 V DC. The maximum pulse speci
fi
cations are 75 W peak, 10
μ
s pulse
width, 0.001 duty cycle, and
≥
30 dB attenuation. For the RSA5100A Series, the maximum voltage rating is ±5 V DC. The
maximum pulse speci
fi
cations are 50 W peak, <10 µs pulse width, 0.001 duty cycle and
≥
30 dB attenuation.